2008K@ Home Blood lines Combiset True Flow Series for Hemodialysis use Catalog Number: 03-2962-3. Intended for use as the extracorporeal blood circuit during Hemodialysis and is indicated for use with conventional and high flux negative pressure hemodialyzer equipment.

Published November 20, 2013 · Recall initiated: August 29, 2013

Summary

FDA records medical device Class II recall Z-0184-2014, published November 20, 2013 and initiated August 29, 2013. Recalling firm: Fresenius Medical Care Holdings, Inc.; Status: Terminated; Reported quantity: 8,528 cases; Lot: All lot codes. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Updated Instructions for Use: Bloodline connections not tightened per instructions in the operator's manual may disconnect and lead to blood loss

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Fresenius Medical Care Holdings, Inc.
Distributor
Worldwide Distribution-USA (nationwide) and the country of Canada.
Lot
All lot codes
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was 2008K@ Home Blood lines Combiset True Flow Series for Hemodialysis use Catalog Number: 03-2962-3. Intended for use as the extracorporeal blood circuit during Hemodialysis and is indicated for use with conventional and high flux negative pressure hemodialyzer equipment. recalled?

Yes. FDA published recall Z-0184-2014 on November 20, 2013. Reported hazard: Updated Instructions for Use: Bloodline connections not tightened per instructions in the operator's manual may disconnect and lead to blood loss

What should I do if I have this product?

Voluntary: Firm initiated · Letter