FDAFDA · Food and drugs·Recall number Z-0184-2014
2008K@ Home Blood lines Combiset True Flow Series for Hemodialysis use Catalog Number: 03-2962-3. Intended for use as the extracorporeal blood circuit during Hemodialysis and is indicated for use with conventional and high flux negative pressure hemodialyzer equipment.
Published November 20, 2013 · Recall initiated: August 29, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Updated Instructions for Use: Bloodline connections not tightened per instructions in the operator's manual may disconnect and lead to blood loss
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Fresenius Medical Care Holdings, Inc.
- Distributor
- Worldwide Distribution-USA (nationwide) and the country of Canada.
- Lot
- All lot codes
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States