Reinforced Dual Lumen Cannula, 32Fr., REF: VV32F, Sterile EO Indications: The OriGen Reinforced Dual Lumen Cannula is indicated for use as a single cannula for both venous drainage and re-infusion of blood in the internal jugular vein during extracorporeal life support procedures.
Published October 24, 2018 · Recall initiated: August 21, 2018
Summary
FDA records medical device Class II recall Z-0184-2019, published October 24, 2018 and initiated August 21, 2018. Recalling firm: OriGen Biomedical, Inc.; Status: Terminated; Reported quantity: 73; Lot: UDI: 00816203022148, Lot: P18806. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
These catheters may have the following deficiencies: 1) barium sulfate omission, and 2) actual catheter tip length is one centimeter shorter than printed on the tip
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- OriGen Biomedical, Inc.
- Distributor
- Worldwide Distribution: U.S.(nationwide) to states of: IN, NY, CA, MA, TX, OH, KY, MI, MS, NM, NC, VA, WI, TN, CT, FL, AR, MO, GA, LA, DC, OR, MN, SC, OK, AZ, UT, RI, SD, WA, PA, AL, MD, IL, HI; and countries (OUS) of: Argentina, Austria, Belgium, CANADA, Colombia, EU, Germany, India, Italy, Japan, Kingdom of Saudi Arabia, Netherlands, New Zealand, Poland, Qatar, Romania, South Africa, SWEDEN, Thailand, UK, United Arab Emirates, and VIETNAM.
- Lot
- UDI: 00816203022148, Lot: P18806
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAccessory Sets Syringes: 6ml syringe with a female luer with 15cm sealed SCD tubing, REF: RF-T15, and 60mlFebruary 21, 2024
- FDAECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV28FJune 12, 2019
- FDAReinforced Dual Lumen Cannula, 16Fr., REF: VV16F, Sterile EO Indications: The OriGen Reinforced Dual LumenOctober 24, 2018
- FDAAtellica Sample Handler Prime (SMN 11069001). The Atellica" Solution is a multi-component system for in vitroOctober 24, 2018
- FDATerumo® Advanced Perfusion System 1 Electronic Patient Gas System, Catalog#: 801188October 24, 2018
- FDACentricity Universal Viewer with Cross Enterprise Display This impacts product versions 6.0 SP1 and higher.October 24, 2018
Questions about this recall
Was Reinforced Dual Lumen Cannula, 32Fr., REF: VV32F, Sterile EO Indications: The OriGen Reinforced Dual Lumen Cannula is indicated for use as a single cannula for both venous drainage and re-infusion of blood in the internal jugular vein during extracorporeal life support procedures. recalled?
Yes. FDA published recall Z-0184-2019 on October 24, 2018. Reported hazard: These catheters may have the following deficiencies: 1) barium sulfate omission, and 2) actual catheter tip length is one centimeter shorter than printed on the tip
What should I do if I have this product?
Voluntary: Firm initiated · Letter