Accessory Sets Syringes: 6ml syringe with a female luer with 15cm sealed SCD tubing, REF: RF-T15, and 60ml syringe on a female luer with 15cm sealed SCD tubing, REF: 15-RF60-T
Published February 21, 2024 · Recall initiated: December 19, 2023
Summary
FDA records medical device Class II recall Z-1068-2024, published February 21, 2024 and initiated December 19, 2023. Recalling firm: OriGen Biomedical, Inc.; Status: Ongoing; Reported quantity: 3613 Cases; Lot: REF/UDI-DI/Lots: RF-T15/10816203020205/ V23277, V23278, V23639; 15-RF60-T/10816203020571/V23281. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Accessory sets contain a syringes not yet qualified for the accessory's intended purpose; due to design changes to the syringe barrels and a polypropylene resin change that resulted in loss of gamma compatibility, for accessory set syringes that were gamma sterilized, which resulted in yellowed syringes.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- OriGen Biomedical, Inc.
- Distributor
- US: FL, WA, NJ, MA, OH, CA, MD, TX, NE, ND OUS: GB, CA,NL,CH, PL, DE
- Lot
- REF/UDI-DI/Lots: RF-T15/10816203020205/ V23277, V23278, V23639; 15-RF60-T/10816203020571/V23281
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAFT (Allograft Filler Tube) Diverted Tube (3/4 Filled), Product Code 227005 (bone void filler)February 21, 2024
- FDACaptus 4000e Thyroid Uptake Systems- Intended to be used by trained Nuclear Medicine Technologists or NuclearFebruary 21, 2024
- FDAAFT (Allograft Filler Tube) Straight Tube (1/2 Filled), Product Code 227010 (bone void filler)February 21, 2024
- FDAECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV28FJune 12, 2019
- FDAReinforced Dual Lumen Cannula, 32Fr., REF: VV32F, Sterile EO Indications: The OriGen Reinforced Dual LumenOctober 24, 2018
- FDAReinforced Dual Lumen Cannula, 16Fr., REF: VV16F, Sterile EO Indications: The OriGen Reinforced Dual LumenOctober 24, 2018
Questions about this recall
Was Accessory Sets Syringes: 6ml syringe with a female luer with 15cm sealed SCD tubing, REF: RF-T15, and 60ml syringe on a female luer with 15cm sealed SCD tubing, REF: 15-RF60-T recalled?
Yes. FDA published recall Z-1068-2024 on February 21, 2024. Reported hazard: Accessory sets contain a syringes not yet qualified for the accessory's intended purpose; due to design changes to the syringe barrels and a polypropylene resin change that resulted in loss of gamma compatibility, for accessory set syringes that were gamma sterilized, which resulted in yellowed syringes.
What should I do if I have this product?
Voluntary: Firm initiated · Letter