Back to search
FDAFDA · Food and drugs·Recall number Z-1068-2024

Accessory Sets Syringes: 6ml syringe with a female luer with 15cm sealed SCD tubing, REF: RF-T15, and 60ml syringe on a female luer with 15cm sealed SCD tubing, REF: 15-RF60-T

Published February 21, 2024 · Recall initiated: December 19, 2023

Summary

FDA records medical device Class II recall Z-1068-2024, published February 21, 2024 and initiated December 19, 2023. Recalling firm: OriGen Biomedical, Inc.; Status: Ongoing; Reported quantity: 3613 Cases; Lot: REF/UDI-DI/Lots: RF-T15/10816203020205/ V23277, V23278, V23639; 15-RF60-T/10816203020571/V23281. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Accessory sets contain a syringes not yet qualified for the accessory's intended purpose; due to design changes to the syringe barrels and a polypropylene resin change that resulted in loss of gamma compatibility, for accessory set syringes that were gamma sterilized, which resulted in yellowed syringes.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
OriGen Biomedical, Inc.
Distributor
US: FL, WA, NJ, MA, OH, CA, MD, TX, NE, ND OUS: GB, CA,NL,CH, PL, DE
Lot
REF/UDI-DI/Lots: RF-T15/10816203020205/ V23277, V23278, V23639; 15-RF60-T/10816203020571/V23281
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Accessory Sets Syringes: 6ml syringe with a female luer with 15cm sealed SCD tubing, REF: RF-T15, and 60ml syringe on a female luer with 15cm sealed SCD tubing, REF: 15-RF60-T recalled?

Yes. FDA published recall Z-1068-2024 on February 21, 2024. Reported hazard: Accessory sets contain a syringes not yet qualified for the accessory's intended purpose; due to design changes to the syringe barrels and a polypropylene resin change that resulted in loss of gamma compatibility, for accessory set syringes that were gamma sterilized, which resulted in yellowed syringes.

What should I do if I have this product?

Voluntary: Firm initiated · Letter