FDAFDA · Food and drugs·Recall number Z-0192-2013
Cliniqa Liquid QCTM Cardiac Marker Control for i-STAT Level 1 (List Number 06F12-09; Lot Number 071039) Cliniqa Liquid QCTM Cardiac Marker Control for i-STAT Level 2 (List Number 06F12-10; Lot Number 081039) Cliniqa Liquid QCTM Cardiac Marker Control for i-STAT Level 3 (List Number 06F12-11; Lot Number 091039) Usage: Assayed human plasma which may be used to verify the integrity of newly received i-STAT CK-MB cartridges.
Published November 14, 2012 · Recall initiated: July 19, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Abbott Point of Care (APOC) has determined that for the Cliniqa Liquid QC Cardiac Marker Control for i-STAT Level 1 (List #06F12-09; Lot Number 071039); i-STAT Level 2 (List #06F12-10; Lot Number 081039); and i-STAT Level 3 (List #06F12-11; Lot Number 091039) results may be generated below the lower value assignment range for some vials.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Abbott Point Of Care Inc.
- Distributor
- Worldwide distribution: USA (nationwide) including states of: AZ, GA, ID, IL, LA, MD, MO, MN, NC, NV, OK, PA, TN, TX, VA and WI; and country of: Italy.
- Lot
- Cliniqa Liquid QC Cardiac Marker Control for i-STAT Level 1 (List #06F12-09; Lot Number 071039) Cliniqa Liquid QC Cardiac Marker Control for i-STAT Level 2 (List #06F12-10; Lot Number 081039) Cliniqa Liquid QC Cardiac Marker Control for i-STAT Level 3 (List #06F12-11; Lot Number 091039)
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States