i-STAT EG6+ cartridge; List Number: 03P77-25;
Published May 6, 2026 · Recall initiated: April 1, 2026
Summary
FDA records medical device Class II recall Z-2000-2026, published May 6, 2026 and initiated April 1, 2026. Recalling firm: Abbott Point Of Care Inc.; Status: Ongoing; Reported quantity: 30775 units (25625 US, 5150 OUS); Lot: List Number: 03P77-25; UDI-DI: 00054749000050; Lot number: N25223A. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Abbott Point Of Care Inc.
- Distributor
- Worldwide distribution - US Nationwide and the countries of Canada, AUSTRIA, BANGLADESH, BELGIUM, FINLAND, FRANCE, GERMANY, IRELAND, ITALY, LUXEMBOURG, MAURITIUS, NETHERLANDS, NORWAY, POLAND, PORTUGAL, SPAIN, SWEDEN, SWITZERLAND, UNITED KINGDOM.
- Lot
- List Number: 03P77-25; UDI-DI: 00054749000050; Lot number: N25223A;
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAi-STAT EG7+ cartridge; List Number: 03P76-25;May 6, 2026
- FDAi-STAT G3+ cartridge; List Number: 03P78-26;May 6, 2026
- FDAHeraeus, PALAMIX duo. Material Number: 66057897.May 6, 2026
- FDAPhilips Bridge Prep Kit REF: K12-09098B Sterile EO, Rx ONLYMay 6, 2026
- FDAi-STAT EG6+ cartridge. List Number: 03P77-25.September 24, 2025
- FDAAbbott i-STAT CG8+ cartridges - IVD Intended for quantitative measurement of sodium, potassium, ionizedJanuary 27, 2021
Questions about this recall
Was i-STAT EG6+ cartridge; List Number: 03P77-25; recalled?
Yes. FDA published recall Z-2000-2026 on May 6, 2026. Reported hazard: Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.
What should I do if I have this product?
Voluntary: Firm initiated · Letter