Back to search
FDAFDA · Food and drugs·Recall number Z-0195-2027

Medtronic NDI Passive Spheres REF: 8801074, Sterile EO, 4/Tray 12 Pk., Rxonly

Published October 28, 2026 · Recall initiated: August 31, 2026

Summary

FDA records medical device Class II recall Z-0195-2027, published October 28, 2026 and initiated August 31, 2026. Recalling firm: Medtronic Navigation, Inc.; Status: Ongoing; Reported quantity: 83,594 units; Lot: UDI: 20805832070023/ Lot codes: 2503141 2503241 2503411 2503491 2504081 2504091 2504211 2504301 2504331 2505211 2505221 2505351 2505641 2505651 2505681 2506081 2506161 2506341 2506551 2506581 2509781 2509791 2509811 2510121 2510741 2510751 2511101 2511121 2511501 2511601. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Sterile Passive spheres and biopsy needle kits contain variation in the reflective material which may affect the ability of the navigation system to detect and track the spheres.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Medtronic Navigation, Inc.
Distributor
US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY OUS: Canary Islands, Iceland, Northern Ireland, Martinique, China , Taiwan, Hong Kong,
Lot
UDI: 20805832070023/ Lot codes: 2503141 2503241 2503411 2503491 2504081 2504091 2504211 2504301 2504331 2505211 2505221 2505351 2505641 2505651 2505681 2506081 2506161 2506341 2506551 2506581 2509781 2509791 2509811 2510121 2510741 2510751 2511101 2511121 2511501 2511601
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Medtronic NDI Passive Spheres REF: 8801074, Sterile EO, 4/Tray 12 Pk., Rxonly recalled?

Yes. FDA published recall Z-0195-2027 on October 28, 2026. Reported hazard: Sterile Passive spheres and biopsy needle kits contain variation in the reflective material which may affect the ability of the navigation system to detect and track the spheres.

What should I do if I have this product?