Back to search
FDAFDA · Food and drugs·Recall number Z-0197-2027

Medtronic Biopsy Needle Kit, REF: 9733068, Passive, Sterile EO, Rxonly

Published October 28, 2026 · Recall initiated: August 31, 2026

Summary

FDA records medical device Class II recall Z-0197-2027, published October 28, 2026 and initiated August 31, 2026. Recalling firm: Medtronic Navigation, Inc.; Status: Ongoing; Reported quantity: 3746 units; Lot: UDI: 00643169702943/ Lots: 066500526 066501226 066503326 066504726 066506326 066509826; UDI:00643169030121/Lots: 066500526 066506326. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Sterile Passive spheres and biopsy needle kits contain variation in the reflective material which may affect the ability of the navigation system to detect and track the spheres.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Medtronic Navigation, Inc.
Distributor
US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY OUS: Canary Islands, Iceland, Northern Ireland, Martinique, China , Taiwan, Hong Kong,
Lot
UDI: 00643169702943/ Lots: 066500526 066501226 066503326 066504726 066506326 066509826; UDI:00643169030121/Lots: 066500526 066506326
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Medtronic Biopsy Needle Kit, REF: 9733068, Passive, Sterile EO, Rxonly recalled?

Yes. FDA published recall Z-0197-2027 on October 28, 2026. Reported hazard: Sterile Passive spheres and biopsy needle kits contain variation in the reflective material which may affect the ability of the navigation system to detect and track the spheres.

What should I do if I have this product?

Voluntary: Firm initiated · Letter