FDAFDA · Food and drugs·Recall number Z-0197-2015
Integra Titan Reverse Shoulder System right and left head cutting templates, 142.5 degrees. The Reverse Shoulder System Cutting Template, 142.5 degrees is an accessory for use with the Integra Titan Reverse Shoulder System. It is used to project the osteotomy angle on the humeral head.
Published November 19, 2014 · Recall initiated: October 24, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A single lot of left and a single lot of right Reverse Shoulder System cutting templates were manufactured incorrectly. Specifically, the threaded handle has been welded backwards on the template resulting in the cutting angle being the reverse of what it should be.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Integra LifeSciences Corp.
- Distributor
- Nationwide Distribution including the states of TX, CA, OH, MS, VA, IL, TN, GA, CA, FL, and AZ.
- Lot
- Lot Numbers PM0278 and PM0277
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States