Integra Titan Reverse Shoulder System right and left head cutting templates, 142.5 degrees. The Reverse Shoulder System Cutting Template, 142.5 degrees is an accessory for use with the Integra Titan Reverse Shoulder System. It is used to project the osteotomy angle on the humeral head.

Published November 19, 2014 · Recall initiated: October 24, 2014

Summary

FDA records medical device Class II recall Z-0197-2015, published November 19, 2014 and initiated October 24, 2014. Recalling firm: Integra LifeSciences Corp.; Status: Terminated; Reported quantity: 12 units; Lot: Lot Numbers PM0278 and PM0277. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A single lot of left and a single lot of right Reverse Shoulder System cutting templates were manufactured incorrectly. Specifically, the threaded handle has been welded backwards on the template resulting in the cutting angle being the reverse of what it should be.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Integra LifeSciences Corp.
Distributor
Nationwide Distribution including the states of TX, CA, OH, MS, VA, IL, TN, GA, CA, FL, and AZ.
Lot
Lot Numbers PM0278 and PM0277
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Integra Titan Reverse Shoulder System right and left head cutting templates, 142.5 degrees. The Reverse Shoulder System Cutting Template, 142.5 degrees is an accessory for use with the Integra Titan Reverse Shoulder System. It is used to project the osteotomy angle on the humeral head. recalled?

Yes. FDA published recall Z-0197-2015 on November 19, 2014. Reported hazard: A single lot of left and a single lot of right Reverse Shoulder System cutting templates were manufactured incorrectly. Specifically, the threaded handle has been welded backwards on the template resulting in the cutting angle being the reverse of what it should be.

What should I do if I have this product?

Voluntary: Firm initiated · Letter