The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation.

Published November 19, 2014 · Recall initiated: October 10, 2014

Summary

FDA records medical device Class II recall Z-0196-2015, published November 19, 2014 and initiated October 10, 2014. Recalling firm: Steris Corporation; Status: Terminated; Reported quantity: 999; Lot: 5085 and 5085SRT. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The 5085 and 5085SRT stainless steel column shrouds may become damaged as a result of misalignment during extreme table articulation, excessive side force on the table column, or pressure on the base of the table. Severe damage to the shrouds may prevent the user from raising or lowering the surgical table.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Steris Corporation
Distributor
US distribution: AL, AZ, AR, CA, CO, CN, DC, FL, GA, HI, ID, IL, IN, IA, KN, KY, LA, ME, MD, MI, MN, MS, MO, NB, NV, NY, NC, ND, OH, OK, OR, PN, RI, SC, SD, TN, TX, VA, WV, WI. Foriegn: Austrialia, Brazil, Canada, China, Ecuador, India, Korea, Mexico, Pnama, and Thailand.
Lot
5085 and 5085SRT
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation. recalled?

Yes. FDA published recall Z-0196-2015 on November 19, 2014. Reported hazard: The 5085 and 5085SRT stainless steel column shrouds may become damaged as a result of misalignment during extreme table articulation, excessive side force on the table column, or pressure on the base of the table. Severe damage to the shrouds may prevent the user from raising or lowering the surgical table.

What should I do if I have this product?

Voluntary: Firm initiated · Letter