The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation.
Published November 19, 2014 · Recall initiated: October 10, 2014
Summary
FDA records medical device Class II recall Z-0196-2015, published November 19, 2014 and initiated October 10, 2014. Recalling firm: Steris Corporation; Status: Terminated; Reported quantity: 999; Lot: 5085 and 5085SRT. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The 5085 and 5085SRT stainless steel column shrouds may become damaged as a result of misalignment during extreme table articulation, excessive side force on the table column, or pressure on the base of the table. Severe damage to the shrouds may prevent the user from raising or lowering the surgical table.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Steris Corporation
- Distributor
- US distribution: AL, AZ, AR, CA, CO, CN, DC, FL, GA, HI, ID, IL, IN, IA, KN, KY, LA, ME, MD, MI, MN, MS, MO, NB, NV, NY, NC, ND, OH, OK, OR, PN, RI, SC, SD, TN, TX, VA, WV, WI. Foriegn: Austrialia, Brazil, Canada, China, Ecuador, India, Korea, Mexico, Pnama, and Thailand.
- Lot
- 5085 and 5085SRT
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020117October 21, 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115October 21, 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020116October 21, 2026
- FDAVERIFY Biological Steam Test Pack Model/Catalog Number: S3069October 14, 2026
- FDASeries OEC 9800. The Series OEC 9800 is designed to provide fluoroscopic and spot-film imaging of theNovember 19, 2014
- FDAIntegra Titan Reverse Shoulder System right and left head cutting templates, 142.5 degrees. The ReverseNovember 19, 2014
Questions about this recall
Was The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation. recalled?
Yes. FDA published recall Z-0196-2015 on November 19, 2014. Reported hazard: The 5085 and 5085SRT stainless steel column shrouds may become damaged as a result of misalignment during extreme table articulation, excessive side force on the table column, or pressure on the base of the table. Severe damage to the shrouds may prevent the user from raising or lowering the surgical table.
What should I do if I have this product?
Voluntary: Firm initiated · Letter