Plum A+ Hyperbaric Infusion Pump; List Number 11005; Labeled in part: PLUM A+ HYPERBARIC; Product Usage: The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care
Published November 20, 2013 · Recall initiated: February 5, 2012
Summary
FDA records medical device Class II recall Z-0203-2014, published November 20, 2013 and initiated February 5, 2012. Recalling firm: Hospira Inc.; Status: Terminated; Reported quantity: List No. 11005: 179 units; Lot: List No. 11005; Serial Numbers: 16792201, 16792202, 16792205, 16792206, 16792208, 16792209, 16792210, 16792211, 16792213, 16792216, 16792217, 16792219, 16792220, 16792222, 16792223, 16792224, 16792225, 16792226, 16792227, 16792228, 16792230, 16792232, 16792233, 16792234, 16792235, 16792236, 16792238, 16792239, 16792240, 16792241, 16792242, 16792244, 16792245, 16792246, 16792248, 16792249, 16792250, 16792251, 16792252, 16792253, 16792254, 16792255, 16792256, 16792257, 16792258, 16792259, 16792260, 16792261, 16792263, 16792264, 16792265, 16792266, 16792268, 16792269, 16792271, 16792272, 16792273, 18807402, 18807405, 18807406, 18807407, 18807408, 18807409, 18807410, 18807413, 18807414, 18807417, 18807418, 18807420, 18807421, 18807423, 18807427, 18807428, 18807429, 18807430, 18807431, 18807432, 18807433, 18807434, 18807439, 18807443, 18807444, 18807446, 18807449, 18807452, 18807453, 18807455, 18807456, 18807459, 18807463, 18807466, 18807472, 18807474, 18807478, 18807484, 18807485, 18807486, 18807490, 18807491, 18807494, 18807496, 18807499, 18807500, 18807501, 18807506, 18807509, 18807510, 18807511, 18807513, 18807514, 18807518, 18807523, 18807524, 18807529, 18807535, 18807536, 18807539, 18807540, 18807541, 18807542, 18807543, 18807545, 18807546, 18807556, 18807557, 18807568, 18807574, 18807575, 18807576, 18807578, 18807581, 18807584, 18807585, 18807586, 18807588, 18807589, 18807590, 18807591, 18807596, 18807597, 18807598, 18807599, 18807600, 18807601, 18807602, 18807604, 18807605, 18807607, 18807608, 18807609, 18807610, 18807611, 18807612, 18807613, 18807614, 18807617, 18807618, 18807619, 18807620, 18807623, 18807625, 18807626, 18807627, 18807628, 18807630, 18807631, 18807632, 18807633, 18807634, 18807636, 18807637, 18807639, 18807640, 18807695, 18807696, 18807697, 18807698, 18807699, 18807700. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Hospira has received reports of distal pressure sensor calibration drift on Plum A+ Infusers. The following error codes may be displayed during setup or infusion signifying possible calibration drift: E180/N180, E181/N181, E186/N186, E187/N187 or E346. Additionally, if the pressure sensor calibration has drifted, users may experience false distal occlusion alarms or distal occlusions that may not
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Hospira Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, District of Columbia, Puerto Rico and U.S. Virgin Islands; and countries of: Croatia, Dominican Republic, Turks and Caicos.
- Lot
- List No. 11005; Serial Numbers: 16792201, 16792202, 16792205, 16792206, 16792208, 16792209, 16792210, 16792211, 16792213, 16792216, 16792217, 16792219, 16792220, 16792222, 16792223, 16792224, 16792225, 16792226, 16792227, 16792228, 16792230, 16792232, 16792233, 16792234, 16792235, 16792236, 16792238, 16792239, 16792240, 16792241, 16792242, 16792244, 16792245, 16792246, 16792248, 16792249, 16792250, 16792251, 16792252, 16792253, 16792254, 16792255, 16792256, 16792257, 16792258, 16792259, 16792260, 16792261, 16792263, 16792264, 16792265, 16792266, 16792268, 16792269, 16792271, 16792272, 16792273, 18807402, 18807405, 18807406, 18807407, 18807408, 18807409, 18807410, 18807413, 18807414, 18807417, 18807418, 18807420, 18807421, 18807423, 18807427, 18807428, 18807429, 18807430, 18807431, 18807432, 18807433, 18807434, 18807439, 18807443, 18807444, 18807446, 18807449, 18807452, 18807453, 18807455, 18807456, 18807459, 18807463, 18807466, 18807472, 18807474, 18807478, 18807484, 18807485, 18807486, 18807490, 18807491, 18807494, 18807496, 18807499, 18807500, 18807501, 18807506, 18807509, 18807510, 18807511, 18807513, 18807514, 18807518, 18807523, 18807524, 18807529, 18807535, 18807536, 18807539, 18807540, 18807541, 18807542, 18807543, 18807545, 18807546, 18807556, 18807557, 18807568, 18807574, 18807575, 18807576, 18807578, 18807581, 18807584, 18807585, 18807586, 18807588, 18807589, 18807590, 18807591, 18807596, 18807597, 18807598, 18807599, 18807600, 18807601, 18807602, 18807604, 18807605, 18807607, 18807608, 18807609, 18807610, 18807611, 18807612, 18807613, 18807614, 18807617, 18807618, 18807619, 18807620, 18807623, 18807625, 18807626, 18807627, 18807628, 18807630, 18807631, 18807632, 18807633, 18807634, 18807636, 18807637, 18807639, 18807640, 18807695, 18807696, 18807697, 18807698, 18807699, 18807700
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAInfant 25% DEXTROSE Injection, USP, 2.5 g (250 mg/mL), 10 mL Unit of Use Single-dose Syringe per carton, RxMay 31, 2017
- FDAMinibore Extension Set, 59 Inch with Spin Lock Collar, Non-DEHP, Latex-Free Intended use: For theJanuary 18, 2017
- FDAFentanyl Citrate Injection, USP, 100 mcg Fentanyl/ 2mL, Single Dose, 50 mcg/mL, Rx only, Hospira, Inc., LakeNovember 30, 2016
- FDASymbiq Two Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids. List numberNovember 16, 2016
- FDASynthes 8.5 MM Medullary Reamer Product Usage: An orthopedic manual surgical instrument is a nonpoweredNovember 20, 2013
- FDAPlum A+ Infusion Pump; List Number: 11971; Product Usage: The Plum A+ Infusion Systems are a multipurposeNovember 20, 2013
Questions about this recall
Was Plum A+ Hyperbaric Infusion Pump; List Number 11005; Labeled in part: PLUM A+ HYPERBARIC; Product Usage: The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care recalled?
Yes. FDA published recall Z-0203-2014 on November 20, 2013. Reported hazard: Hospira has received reports of distal pressure sensor calibration drift on Plum A+ Infusers. The following error codes may be displayed during setup or infusion signifying possible calibration drift: E180/N180, E181/N181, E186/N186, E187/N187 or E346. Additionally, if the pressure sensor calibration has drifted, users may experience false distal occlusion alarms or distal occlusions that may not
What should I do if I have this product?
Voluntary: Firm initiated · Letter