Fisher and Paykel Healthcare (FPH) IW900-Series Infant Warmer Fisher and Paykel; Healthcare CosyCot and Kit Replacement Head Case Assembly It is designed to assist with the thermoregulation of infants weighing up to 10kg and is used by labor and delivery rooms and pediatric intensive care units. Infant Warmers Model Number (REF) IW910XXX, IW920XXX, IW931XXX, IW932XXX, IW933XXX, IW934XXX, IW951XXX, IW952XXX, IW953XXX, IW954XXX, IW980XXX, IW990XXX Spare Parts Model Number (REF) 043041129, 043041130, 043041131, 043042359, 648040142

Published November 19, 2014 · Recall initiated: November 6, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Breakage of the nut that secures the heater head in place, which can cause the heater head to become partially detached and swing towards the bassinet. There is the potential that a heater head could contact a patient situated on the bassinet and cause a serious injury.

What to do

Voluntary: Firm initiated · Telephone

Recall details

Manufacturer
Fisher & Paykel Healthcare, Ltd.
Distributor
Worldwide Distribution - USA (nationwide) and Internationally to Albania, Algeria, Australia, Austria, Bahrain, Bangladesh, Bekgium, Bermuda, Chile, China, Croatia, Dominican Republic, Estonia, Finland, France, Germany, Great Britain, Greece, Hong Kong, Iceland, India, Indonesia, Iran, Ireland, Israel, Jordan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Mauritius, Mexico. Morocco, New Zealand, Nigeria, Northern Ireland, Norway, Pakistan, Palestinian Territory, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, singapore, Slovenia, South Africa, Spain, Sri Lanka, Sultanate of Oman, Syria, Taiwan, Thailand, The Netherlands, Tunisia, Turkey, United Arab Emirates, Uruguay, Venezuela, Vietnam, and West Bank.
Lot
Affected Serial (SN) and Lot (LOT) Numbers Infant Warmers: 071116ZZZZZZ - 090810ZZZZZZ Spare Parts: 071116 - 090810
Risk classification
Class II
Area (FDA)
device
Country
New Zealand