FDAFDA · Food and drugs·Recall number Z-0228-2025

a. Hillrom Centrella Smart+ Medical-Surgical Hospital Bed, Product codes: P7900B000001, P7900B100010, P7900B100024, P7900B100208, P7900B100273, P7900B100304, P7900B000005, P7900B100011, P7900B100025, P7900B100212, P7900B100277, P7900B100307, P7900B000010, P7900B100017, P7900B100026, P7900B100222, P7900B100281, P7900B1SPL05, P7900B000011, P7900B100018, P7900B100028, P7900B100223, P7900B100287, P7900B1SPL06, P7900B000020, P7900B100019, P7900B100039, P7900B100224, P7900B100288, P7900B000022, P7900B100020, P7900B100090, P7900B100231, P7900B100292, P7900B100001, P7900B100021, P7900B100161, P7900B100237, P7900B100293, P7900B100005, P7900B100022, P7900B100203, P7900B100242, P7900B100294, P7900B100009, P7900B100023, P7900B100205, P7900B100264, and P7900B100296. b. Hillrom Centrella Smart+ Medical-Surgical Hospital Bed, Product codes: P7900A000001, P7900A000010, P7900A000019, P7900A000039, P7900A000005, P7900A000011, P7900A000022, and P7900A000063.

Published November 6, 2024 · Recall initiated: September 18, 2024

Summary

FDA records medical device Class II recall Z-0228-2025, published November 6, 2024 and initiated September 18, 2024. Recalling firm: Baxter Healthcare Corporation; Status: Ongoing; Reported quantity: 2,967 beds; Lot: All serial numbers with software version 1.27.000. UDI-DI 00887761985162: Product codes P7900B000001, P7900B100010, P7900B100024, P7900B100208, P7900B100273, P7900B100304, P7900B000005, P7900B100011, P7900B100025, P7900B100212, P7900B100277, P7900B100307, P7900B000010, P7900B100017, P7900B100026, P7900B100222, P7900B100281, P7900B1SPL05, P7900B000011, P7900B100018, P7900B100028, P7900B100223, P7900B100287, P7900B1SPL06, P7900B000020, P7900B100019, P7900B100039, P7900B100224, P7900B100288, P7900B000022, P7900B100020, P7900B100090, P7900B100231, P7900B100292, P7900B100001, P7900B100021, P7900B100161, P7900B100237, P7900B100293, P7900B100005, P7900B100022, P7900B100203, P7900B100242, P7900B100294, P7900B100009, P7900B100023, P7900B100205, P7900B100264, and P7900B100296. UDI-DI 00887761999961: Product codes P7900A000001, P7900A000010, P7900A000019, P7900A000039, P7900A000005, P7900A000011, P7900A000022, and P7900A000063. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The beds have a potential for an interface issue with the Rauland Responder 5 Nurse Call System where the bed exit system will alert at the bed, but may fail to send a remote alert through the nurse call system.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Baxter Healthcare Corporation
Distributor
Worldwide distribution - US Nationwide. There was government and military distribution. The countries of Canada and Australia.
Lot
All serial numbers with software version 1.27.000. UDI-DI 00887761985162: Product codes P7900B000001, P7900B100010, P7900B100024, P7900B100208, P7900B100273, P7900B100304, P7900B000005, P7900B100011, P7900B100025, P7900B100212, P7900B100277, P7900B100307, P7900B000010, P7900B100017, P7900B100026, P7900B100222, P7900B100281, P7900B1SPL05, P7900B000011, P7900B100018, P7900B100028, P7900B100223, P7900B100287, P7900B1SPL06, P7900B000020, P7900B100019, P7900B100039, P7900B100224, P7900B100288, P7900B000022, P7900B100020, P7900B100090, P7900B100231, P7900B100292, P7900B100001, P7900B100021, P7900B100161, P7900B100237, P7900B100293, P7900B100005, P7900B100022, P7900B100203, P7900B100242, P7900B100294, P7900B100009, P7900B100023, P7900B100205, P7900B100264, and P7900B100296. UDI-DI 00887761999961: Product codes P7900A000001, P7900A000010, P7900A000019, P7900A000039, P7900A000005, P7900A000011, P7900A000022, and P7900A000063.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was a. Hillrom Centrella Smart+ Medical-Surgical Hospital Bed, Product codes: P7900B000001, P7900B100010, P7900B100024, P7900B100208, P7900B100273, P7900B100304, P7900B000005, P7900B100011, P7900B100025, P7900B100212, P7900B100277, P7900B100307, P7900B000010, P7900B100017, P7900B100026, P7900B100222, P7900B100281, P7900B1SPL05, P7900B000011, P7900B100018, P7900B100028, P7900B100223, P7900B100287, P7900B1SPL06, P7900B000020, P7900B100019, P7900B100039, P7900B100224, P7900B100288, P7900B000022, P7900B100020, P7900B100090, P7900B100231, P7900B100292, P7900B100001, P7900B100021, P7900B100161, P7900B100237, P7900B100293, P7900B100005, P7900B100022, P7900B100203, P7900B100242, P7900B100294, P7900B100009, P7900B100023, P7900B100205, P7900B100264, and P7900B100296. b. Hillrom Centrella Smart+ Medical-Surgical Hospital Bed, Product codes: P7900A000001, P7900A000010, P7900A000019, P7900A000039, P7900A000005, P7900A000011, P7900A000022, and P7900A000063. recalled?

Yes. FDA published recall Z-0228-2025 on November 6, 2024. Reported hazard: The beds have a potential for an interface issue with the Rauland Responder 5 Nurse Call System where the bed exit system will alert at the bed, but may fail to send a remote alert through the nurse call system.

What should I do if I have this product?

Voluntary: Firm initiated · Letter