Quest Spectrum 3 Fluorescent Imaging System Model/Catalog Number: ART-SYS-0029, ART-SYS-0030 The Quest Spectrum3 is a fluorescent imaging system used in capturing and viewing fluorescent images for the visual assessment of blood flow for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used during plastic and reconstructive surgeries.
Published November 6, 2024 · Recall initiated: September 24, 2024
Summary
FDA records medical device Class II recall Z-0232-2025, published November 6, 2024 and initiated September 24, 2024. Recalling firm: Olympus Corporation of the Americas; Status: Ongoing; Reported quantity: 3 units; Lot: UDI-DI: 08719324487047 Lot Code: Model: ART-SYS-0029, Serial: ART-SYS-2022-F04, ART-SYS-2022-A06, ART-SYS-2022-F05; Model: ART-SYS-0030; No distribution in the USA. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Does not meet the requirements for electromagnetic compatibility (EMC) of the standard IEC 60601-1-2. Electromagnetic interference occurring during a procedure may cause the device to malfunction requiring the replacement of the device. This could potentially result in a prolonged procedure and/or require additional medical intervention
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Olympus Corporation of the Americas
- Distributor
- OH
- Lot
- UDI-DI: 08719324487047 Lot Code: Model: ART-SYS-0029, Serial: ART-SYS-2022-F04, ART-SYS-2022-A06, ART-SYS-2022-F05; Model: ART-SYS-0030; No distribution in the USA
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOlympus M3 Gold Autoclavable Lateral Telescope 70. Model Number: M3-70A.October 7, 2026
- FDAOlympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.October 7, 2026
- FDAOlympus Tripp-Lite P580-025 25ft DisplayPort Cable, Model Number P580-025, an accessory used in conjunctionOctober 7, 2026
- FDAOlympus BiCOAG Hemostasis Probe 10Fr Coaxial Connector. Model/Catalog Number: CD-B610LA.September 30, 2026
- FDAClariTEE Probe- Miniature transesophageal echo ultrasound probe for use with ImaCor ultrasound diagnosticNovember 6, 2024
- FDAAbbott Alinity m HPV AMP Kit, used with the Alinity m System, product codes: a) REF 09N15-095; b) REFNovember 6, 2024
Questions about this recall
Was Quest Spectrum 3 Fluorescent Imaging System Model/Catalog Number: ART-SYS-0029, ART-SYS-0030 The Quest Spectrum3 is a fluorescent imaging system used in capturing and viewing fluorescent images for the visual assessment of blood flow for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used during plastic and reconstructive surgeries. recalled?
Yes. FDA published recall Z-0232-2025 on November 6, 2024. Reported hazard: Does not meet the requirements for electromagnetic compatibility (EMC) of the standard IEC 60601-1-2. Electromagnetic interference occurring during a procedure may cause the device to malfunction requiring the replacement of the device. This could potentially result in a prolonged procedure and/or require additional medical intervention
What should I do if I have this product?
Voluntary: Firm initiated · Letter