Back to search
FDAFDA · Food and drugs·Recall number Z-3299-2026

Olympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.

Published October 7, 2026 · Recall initiated: August 24, 2026

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in the Japanese package insert.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
Olympus Corporation of the Americas
Distributor
International distribution to the country of Japan.
Lot
Model Number: M3-30A. UDI-DI: 00821925005679. All Serial Numbers.
Risk classification
Class II
Area (FDA)
device
Country
United States