AMS AdVance" Male Sling System 720088-01 AMS AdVance" Male Sling System (US) 720088-02 AMS AdVance" Male Sling System (Intl) 72404224 AMS Male TO Sling System (Intl). The AMS AdVance Male Sling System is intended for the placement of a suburethral sling for the treatment of male stress urinary incontinence (SUI).
Published November 14, 2012 · Recall initiated: October 19, 2012
Summary
FDA records medical device Class II recall Z-0259-2013, published November 14, 2012 and initiated October 19, 2012. Recalling firm: American Medical Systems, Inc.; Status: Terminated; Reported quantity: 4872; Lot: Part # Lot/Serial Range 720088-01 722022001 to 796127002 720088-02 722024001 to 790266012 72404224 742528001 to 795689010. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
American Medical Systems® (AMS) is issuing a voluntary recall on a number of lots of the AdVance® Male Sling System (AdVance). AMS has recently received a small number of complaints regarding the resorbable suture in the AdVance sling not lasting for the entire 15 month shelf life of the product. An internal investigation has been conducted and confirmed that some sutures degraded prior to the
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- American Medical Systems, Inc.
- Distributor
- Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the countries of Austria, Australia, Brazil, Canada, Germany, Estonia, Spain, Finland, France, Great Britain, Hong Kong, Croatia, India, Morocco, Mexico, Malaysia, Netherlands, New Zealand, Panama, Poland, Russian Federation, Solomon Islands, Singapore, Thailand, Turkey, Venezuela, South Africa, Chile, Colombia, Costa Rica, Slovenia, Argentina, China, and Taiwan.
- Lot
- Part # Lot/Serial Range 720088-01 722022001 to 796127002 720088-02 722024001 to 790266012 72404224 742528001 to 795689010
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAdVance"Male Sling System, REF 720088-01 & AdVance" XP Male Sling System REF 720163-01 (not commerciallyDecember 24, 2014
- FDAAMS Monarc + Subfascial Hammock with Tensioning Suture REF 72404193 and AMS Monarc C Subfascial Hammock withDecember 24, 2014
- FDAAMS 700 Series Inflatable Penile Prosthesis Accessory Kit (AMS 700 Line Accessory Kit), 72401850. KeithNovember 19, 2014
- FDAAMS 800® Urinary Control System. The AMS 800® Urinary Control System is used to treat urinary incontinence dueMay 22, 2013
- FDAMicroCool Breathable Gown with Towels, X-Larg, X-Long, Sterile The product is a sterile, single use, surgicalNovember 14, 2012
- FDAR-30H , Diagnostic X-Ray Beam-Limiting Device. In combination with any of below X-ray tubes 0.6/1.2P364DK-85,November 14, 2012
Questions about this recall
Was AMS AdVance" Male Sling System 720088-01 AMS AdVance" Male Sling System (US) 720088-02 AMS AdVance" Male Sling System (Intl) 72404224 AMS Male TO Sling System (Intl). The AMS AdVance Male Sling System is intended for the placement of a suburethral sling for the treatment of male stress urinary incontinence (SUI). recalled?
Yes. FDA published recall Z-0259-2013 on November 14, 2012. Reported hazard: American Medical Systems® (AMS) is issuing a voluntary recall on a number of lots of the AdVance® Male Sling System (AdVance). AMS has recently received a small number of complaints regarding the resorbable suture in the AdVance sling not lasting for the entire 15 month shelf life of the product. An internal investigation has been conducted and confirmed that some sutures degraded prior to the
What should I do if I have this product?
Voluntary: Firm initiated · Letter