R-30H , Diagnostic X-Ray Beam-Limiting Device. In combination with any of below X-ray tubes 0.6/1.2P364DK-85, 0.6/1.2P324DK-125, 0.6/1.2P324DK-85, 0.6/1.2P38DE-85 Device Model # R-30H. Subsequent Product Code: IZX Product Usage: The device is the beam-limiting device for X-ray Radiography.
Published November 14, 2012 · Recall initiated: October 28, 2011
Summary
FDA records medical device Class II recall Z-0260-2013, published November 14, 2012 and initiated October 28, 2011. Recalling firm: Shimadzu Medical Systems; Status: Terminated; Reported quantity: 665 units; Lot: Device Model #R-30H. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1) Half-value layer (HVL), which was revised in 2006. Certain devices might not comply with the requirements for them in some x-ray generating condition after the revision in 2006.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Shimadzu Medical Systems
- Distributor
- US Nationwide Distribution
- Lot
- Device Model #R-30H
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADigital Angiography System Trinias:UD150G-40 / D150GC-40; Digital Angiography System BRANSIST safire:UD150G-40June 10, 2026
- FDAMODEL: X-RAY TV SYSTEM SONIALVISION G4April 13, 2022
- FDAMODEL: X-RAY TV SYSSTEM SONIALVISION safire17April 13, 2022
- FDAShimadzu X-ray High Voltage Generator Model No. UD150B-40 on SONIALVISION safire II/17 systems withJanuary 4, 2017
- FDAAMS AdVance" Male Sling System 720088-01 AMS AdVance" Male Sling System (US) 720088-02 AMS AdVance" Male SlingNovember 14, 2012
- FDAR-20J , Diagnostic X-Ray Beam-Limiting Device. In combination with any of below X-Ray tubes: 06/1.2P364DK-85,November 14, 2012
Questions about this recall
Was R-30H , Diagnostic X-Ray Beam-Limiting Device. In combination with any of below X-ray tubes 0.6/1.2P364DK-85, 0.6/1.2P324DK-125, 0.6/1.2P324DK-85, 0.6/1.2P38DE-85 Device Model # R-30H. Subsequent Product Code: IZX Product Usage: The device is the beam-limiting device for X-ray Radiography. recalled?
Yes. FDA published recall Z-0260-2013 on November 14, 2012. Reported hazard: Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1) Half-value layer (HVL), which was revised in 2006. Certain devices might not comply with the requirements for them in some x-ray generating condition after the revision in 2006.
What should I do if I have this product?
Voluntary: Firm initiated · Letter