HX2 Temp Manage SYS 120V The Terumo HX2 Temperature Management System is a selfcontained, dual channel system for supplying temperature controlled water to heat exchangers used in an extracorporeal circuit.
Published November 21, 2012 · Recall initiated: October 16, 2012
Summary
FDA records medical device Class II recall Z-0278-2013, published November 21, 2012 and initiated October 16, 2012. Recalling firm: Terumo Cardiovascular Systems Corporation; Status: Terminated; Reported quantity: 19 units; Lot: Catalog number: 809810 and serial numbers: 1041, 1083, 1084, and 1101-1116. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Based on a retrospective review of quality data, AC Control boards used in the manufacture of the HX2 Temperature Management System were identified as being manufactured using the incorrect flux material. Use of the incorrect flux may contribute to premature board failure. A search of TCVS complaints revealed 3 complaints received related to the HX2 failure. Each failure occurred during set up
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Terumo Cardiovascular Systems Corporation
- Distributor
- Nationwide Distribution including CA, HI, KS, LA, MI, MO, OR, PA, UT, and WI
- Lot
- Catalog number: 809810 and serial numbers: 1041, 1083, 1084, and 1101-1116.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPMJuly 2, 2025
- FDACardiovascular Procedure Kits that include High Crack Check Valve PN 205307A Part Number: 6375 16010 62974April 10, 2024
- FDA(1) Capiox FX25 (West) Advance Oxygenator- with left port, 4-liter Reservoir, XC modified to Advance, SterileDecember 23, 2020
- FDATerumo Capiox NX19 Oxygenator With Reservoir (East Orientation) P/N: 3CX*NX19REAugust 7, 2019
- FDASynthes External Fixation System Tube to Tube Clamp, Part Number 390.007. Intended for use to provideNovember 21, 2012
- FDATested AC Control BD, 120V, X2 The Terumo HX2 Temperature Management System is a selfcontained, dual channelNovember 21, 2012
Questions about this recall
Was HX2 Temp Manage SYS 120V The Terumo HX2 Temperature Management System is a selfcontained, dual channel system for supplying temperature controlled water to heat exchangers used in an extracorporeal circuit. recalled?
Yes. FDA published recall Z-0278-2013 on November 21, 2012. Reported hazard: Based on a retrospective review of quality data, AC Control boards used in the manufacture of the HX2 Temperature Management System were identified as being manufactured using the incorrect flux material. Use of the incorrect flux may contribute to premature board failure. A search of TCVS complaints revealed 3 complaints received related to the HX2 failure. Each failure occurred during set up
What should I do if I have this product?
Voluntary: Firm initiated · Letter