Ray Station Radiation Therapy Treatment Planning System, Model Numbers 3.0, 3.5 and 4.0
Published November 20, 2013 · Recall initiated: October 23, 2013
Summary
FDA records medical device Class II recall Z-0281-2014, published November 20, 2013 and initiated October 23, 2013. Recalling firm: RAYSEARCH LABORATORIES AB; Status: Terminated; Reported quantity: 195 units domestically; Lot: Software version and build numbers 3.0.0.251, 3.5.0.16, 3.5.1.6, 4.0.0.14, 4.0.1.4. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Ray Search Laboratories found an issue with the dose calculation for Elekta motorized wedges in Ray Station 3.0, 3.5 and 4.0. The error presents when a wedge is used for a beam and the wedge angle differs from 60 degrees.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- RAYSEARCH LABORATORIES AB
- Distributor
- US Distribution in the states of: California, Colorado, Delaware, Florida, Maryland, Missouri, New Jersey, New York, Ohio, Pennsylvania, Tennessee, and Virginia.
- Lot
- Software version and build numbers 3.0.0.251, 3.5.0.16, 3.5.1.6, 4.0.0.14, 4.0.1.4.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Sweden
Related recalls
- FDARayCare 2024A SP4. Oncology Information Systems.July 8, 2026
- FDARayCare 2024A SP1. Oncology Information Systems.July 8, 2026
- FDABrand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 14.0.0, 15.0.0, 15.1.3 SoftwareJanuary 28, 2026
- FDABrand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 17.0.0, 17.0.1 Software Version:January 28, 2026
- FDAAcetabular Pressurizer; single use and six units per box. The Acetabular Pressurizer 0206-522-000 is anNovember 20, 2013
- FDADraeger Movita 603x/DVE 803x Ceiling Supply Unit. It is intended for use in operating theaters and intensiveNovember 20, 2013
Questions about this recall
Was Ray Station Radiation Therapy Treatment Planning System, Model Numbers 3.0, 3.5 and 4.0 recalled?
Yes. FDA published recall Z-0281-2014 on November 20, 2013. Reported hazard: Ray Search Laboratories found an issue with the dose calculation for Elekta motorized wedges in Ray Station 3.0, 3.5 and 4.0. The error presents when a wedge is used for a beam and the wedge angle differs from 60 degrees.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail