Ray Station Radiation Therapy Treatment Planning System, Model Numbers 3.0, 3.5 and 4.0

Published November 20, 2013 · Recall initiated: October 23, 2013

Summary

FDA records medical device Class II recall Z-0281-2014, published November 20, 2013 and initiated October 23, 2013. Recalling firm: RAYSEARCH LABORATORIES AB; Status: Terminated; Reported quantity: 195 units domestically; Lot: Software version and build numbers 3.0.0.251, 3.5.0.16, 3.5.1.6, 4.0.0.14, 4.0.1.4. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Ray Search Laboratories found an issue with the dose calculation for Elekta motorized wedges in Ray Station 3.0, 3.5 and 4.0. The error presents when a wedge is used for a beam and the wedge angle differs from 60 degrees.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
RAYSEARCH LABORATORIES AB
Distributor
US Distribution in the states of: California, Colorado, Delaware, Florida, Maryland, Missouri, New Jersey, New York, Ohio, Pennsylvania, Tennessee, and Virginia.
Lot
Software version and build numbers 3.0.0.251, 3.5.0.16, 3.5.1.6, 4.0.0.14, 4.0.1.4.
Risk classification
Class II
Area (FDA)
device
Country
Sweden

Related recalls

Questions about this recall

Was Ray Station Radiation Therapy Treatment Planning System, Model Numbers 3.0, 3.5 and 4.0 recalled?

Yes. FDA published recall Z-0281-2014 on November 20, 2013. Reported hazard: Ray Search Laboratories found an issue with the dose calculation for Elekta motorized wedges in Ray Station 3.0, 3.5 and 4.0. The error presents when a wedge is used for a beam and the wedge angle differs from 60 degrees.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail