FDAFDA · Food and drugs·Recall number Z-2646-2026
RayCare 2024A SP1. Oncology Information Systems.
Published July 8, 2026 · Recall initiated: May 20, 2026
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Beam set delivery note issues in which there is potential data loss if the beam set delivery note is modified simultaneously and/or if a note is updated during a treatment session, the user is not notified of the change.
What to do
Voluntary: Firm initiated
Recall details
- Manufacturer
- RAYSEARCH LABORATORIES AB
- Distributor
- Worldwide - US Nationwide distribution.
- Lot
- Software Version 8.0.1.60193, UDI 0735000201083920240906, Expiration Date: 2030-01-24
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Sweden
Sell products like this?