Liquid Protein Calibrators (SP CAL (LIQ)) Catalogue Number: IT2691 GTIN: 05055273204032 Product Usage: Liquid Protein Calibrators are intended for in vitro diagnostic use in the calibration of ASO, Complement C3, Complement C4, CRP, Ferritin, Haptoglobin, IgA, IgG, IgM, Prealbumin, and Transferrin assays on Clinical Chemistry and Immunoassay systems, Abbott Spectrum, Abbott Aeroset, Abbott Architect i2000, Architect i2000sr, Ace analyser, Bayer Advia 1650, Advia 2400, Advia 1200, Dade Dimension RXL, Dimension AR, Hitachi 704, Hitachi 717, Hitachi 911, Hitachi 917, Hitachi 912, Hitachi 747, Kone progress, AU800, AU600, AU400, AU2700, AU5400, Selectra Vitalab, Synchron CX4, Synchron CX5, Synchron CX7, Synchron LX20, ILAB300, ILAB900, ILAB1800, ILAB600, RX Daytona, RX Imola, Cobas Mira, Cobas Mira S, Cobas Mira Plus systems

Published November 13, 2019 · Recall initiated: August 9, 2019

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The assigned value for Rheumatoid Factor over recovers against the reference material RF serum, 1st British standard from NIBSC. Ref = 64/002. The calibrator values have been adjusted to re-align to the reference material.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Randox Laboratories Ltd.
Distributor
US Nationwide distribution in the states of WV, ME, NJ, CA, and Puerto Rico.
Lot
Batch/Lot # 440973 /1748 1752IT 451414 /1748 1752IT 470531 /1813 1817IT 480859 /1813 1817IT
Risk classification
Class II
Area (FDA)
device
Country
Ireland