Liquid Protein Calibrators (SP CAL (LIQ)) Catalogue Number: IT2691 GTIN: 05055273204032 Product Usage: Liquid Protein Calibrators are intended for in vitro diagnostic use in the calibration of ASO, Complement C3, Complement C4, CRP, Ferritin, Haptoglobin, IgA, IgG, IgM, Prealbumin, and Transferrin assays on Clinical Chemistry and Immunoassay systems, Abbott Spectrum, Abbott Aeroset, Abbott Architect i2000, Architect i2000sr, Ace analyser, Bayer Advia 1650, Advia 2400, Advia 1200, Dade Dimension RXL, Dimension AR, Hitachi 704, Hitachi 717, Hitachi 911, Hitachi 917, Hitachi 912, Hitachi 747, Kone progress, AU800, AU600, AU400, AU2700, AU5400, Selectra Vitalab, Synchron CX4, Synchron CX5, Synchron CX7, Synchron LX20, ILAB300, ILAB900, ILAB1800, ILAB600, RX Daytona, RX Imola, Cobas Mira, Cobas Mira S, Cobas Mira Plus systems
Published November 13, 2019 · Recall initiated: August 9, 2019
Summary
FDA records medical device Class II recall Z-0283-2020, published November 13, 2019 and initiated August 9, 2019. Recalling firm: Randox Laboratories Ltd.; Status: Terminated; Reported quantity: 100 kits; Lot: Batch/Lot # 440973 /1748 1752IT 451414 /1748 1752IT 470531 /1813 1817IT 480859 /1813 1817IT. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The assigned value for Rheumatoid Factor over recovers against the reference material RF serum, 1st British standard from NIBSC. Ref = 64/002. The calibrator values have been adjusted to re-align to the reference material.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Randox Laboratories Ltd.
- Distributor
- US Nationwide distribution in the states of WV, ME, NJ, CA, and Puerto Rico.
- UPC
- 05055273204032
- Lot
- Batch/Lot # 440973 /1748 1752IT 451414 /1748 1752IT 470531 /1813 1817IT 480859 /1813 1817IT
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Ireland
Related recalls
- FDARX Series CYSC- Cystatin C Reagent -Intended for in vitro diagnostic (IVD) use as Randox Cystatin C assay onJuly 31, 2024
- FDAEvidence MultiSTAT with software v 3.7-Analyzer intended for the qualitative determination of parent drugMay 15, 2024
- FDAClinical Chemistry Calibration Serum (Catalog number Cal2351) that is 20x5ml lyophilized human based serumMay 15, 2024
- FDAMicroalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567April 24, 2024
- FDAPrismaflex Control Unit, software versions below 7.21November 13, 2019
- FDAArgon Medical Devices TRI-JECT, 3 Way IV Administration Organizer without Stopcock, REF 041230001A, 25/box,November 13, 2019
Questions about this recall
Was Liquid Protein Calibrators (SP CAL (LIQ)) Catalogue Number: IT2691 GTIN: 05055273204032 Product Usage: Liquid Protein Calibrators are intended for in vitro diagnostic use in the calibration of ASO, Complement C3, Complement C4, CRP, Ferritin, Haptoglobin, IgA, IgG, IgM, Prealbumin, and Transferrin assays on Clinical Chemistry and Immunoassay systems, Abbott Spectrum, Abbott Aeroset, Abbott Architect i2000, Architect i2000sr, Ace analyser, Bayer Advia 1650, Advia 2400, Advia 1200, Dade Dimension RXL, Dimension AR, Hitachi 704, Hitachi 717, Hitachi 911, Hitachi 917, Hitachi 912, Hitachi 747, Kone progress, AU800, AU600, AU400, AU2700, AU5400, Selectra Vitalab, Synchron CX4, Synchron CX5, Synchron CX7, Synchron LX20, ILAB300, ILAB900, ILAB1800, ILAB600, RX Daytona, RX Imola, Cobas Mira, Cobas Mira S, Cobas Mira Plus systems recalled?
Yes. FDA published recall Z-0283-2020 on November 13, 2019. Reported hazard: The assigned value for Rheumatoid Factor over recovers against the reference material RF serum, 1st British standard from NIBSC. Ref = 64/002. The calibrator values have been adjusted to re-align to the reference material.
What should I do if I have this product?
Voluntary: Firm initiated · Letter
How do I know if my unit is part of the recall?
Compare your product details with the notice. UPC: 05055273204032.