Cartesion Prime, Digital PET-CT, Model: PCD-1000A/3D.004, PCD-1000A/3D, is a diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.
Published November 13, 2024 · Recall initiated: September 17, 2024
Summary
FDA records medical device Class II recall Z-0284-2025, published November 13, 2024 and initiated September 17, 2024. Recalling firm: Canon Medical System, USA, INC.; Status: Ongoing; Reported quantity: 27; Lot: UDI-DI: 04987670103810, Software: V10.8 SP0013, V10.15 SP0006. Serial Numbers: 3DA1982001, 3DA2072002, 3DA2072003, 3DA20X2004, 3DA20Y2005, 3DA2112006, 3DA2132007, 3DA2162008, 3DA2192009, 3DA2242010, 3DA2252011, 3DA2262012, 3DA2272013, 3DA2272014, 3DA2282015, 3DA2292016, 3DA22Y2017, 3DA22Z2018, 3DA2322019, 3DA2322020, 3DA2332021, 3DA2332022, 3DA2342023, 3DA2352024, 3DA2352025, 3DB2362026, 3DB2422029. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
When PET-CT system is executing reconstruction, if PET acquisition for another patient is performed (or PET reconstruction for another patient is performed from raw data processing), PET reconstruction control processing may hang up and PET reconstruction may stop as a result, which may necessitate repeat PET/CT examinations.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Canon Medical System, USA, INC.
- Distributor
- US: NV, AR, MN, IA, IL, NC, LA, KS, MS, CA, AL, MD, CO, IN, TN, AZ, WI, VA
- Lot
- UDI-DI: 04987670103810, Software: V10.8 SP0013, V10.15 SP0006. Serial Numbers: 3DA1982001, 3DA2072002, 3DA2072003, 3DA20X2004, 3DA20Y2005, 3DA2112006, 3DA2132007, 3DA2162008, 3DA2192009, 3DA2242010, 3DA2252011, 3DA2262012, 3DA2272013, 3DA2272014, 3DA2282015, 3DA2292016, 3DA22Y2017, 3DA22Z2018, 3DA2322019, 3DA2322020, 3DA2332021, 3DA2332022, 3DA2342023, 3DA2352024, 3DA2352025, 3DB2362026, 3DB2422029
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000FNovember 4, 2026
- FDAALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000VNovember 4, 2026
- FDAALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000CNovember 4, 2026
- FDAAlphenix INFX-8000C/B5October 14, 2026
- FDAElekta Disposable Biopsy Needle Kit for Leksell Stereotactic System, REF 911933, Diameter / 14 Gauge,November 13, 2024
- FDASaphena Medical Venapax XL- Endoscopic Vessel Harvesting System for use in minimally invasive surgery allowingNovember 13, 2024
Questions about this recall
Was Cartesion Prime, Digital PET-CT, Model: PCD-1000A/3D.004, PCD-1000A/3D, is a diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. recalled?
Yes. FDA published recall Z-0284-2025 on November 13, 2024. Reported hazard: When PET-CT system is executing reconstruction, if PET acquisition for another patient is performed (or PET reconstruction for another patient is performed from raw data processing), PET reconstruction control processing may hang up and PET reconstruction may stop as a result, which may necessitate repeat PET/CT examinations.
What should I do if I have this product?
Voluntary: Firm initiated · Letter