FDAFDA · Food and drugs·Recall number Z-0227-2025

Elekta Disposable Biopsy Needle Kit for Leksell Stereotactic System, REF 911933, Diameter / 14 Gauge, containing 6 inner boxes labeled REF 911761, Disposable Biopsy Needle Kit for Leksell Stereotactic System. Each inner box of REF 911761 contains a Tyvek bag labeled REF 912465, Disposable Biopsy Needle and Aspirator Tube. The system is used for for localization and diagnosis of intracranial disorders and their surgical treatment including radiosurgery and stereotactic radiation therapy.

Published November 13, 2024 · Recall initiated: September 25, 2024

Summary

FDA records medical device Class I recall Z-0227-2025, published November 13, 2024 and initiated September 25, 2024. Recalling firm: Elekta Instrument AB; Status: Ongoing; Reported quantity: 111 kits (666 needles); Lot: Batch Number 873874875R Package including 6 needles: UDI: (01)0 7340048 30034 6 (17) 260301 (10) 873874875R (240) 911933 Package including 1 needle: UDI: (01)0 7340048 30800 7 (17) 260301 (10) 873874875R (240) 911761 Inner sterile bag including 1 needle: UDI: (01)0 7340048 30801 4 (17) 260301 (10) 873874875R (240) 912465. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Elekta has become aware that Disposable Biopsy Needles can contain some microscopic debris on the inside of the biopsy needle. The material in the debris is stainless steel, same material as the biopsy needle. No debris has been found on the outer parts of the biopsy needles.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Elekta Instrument AB
Distributor
US: AZ, TX, FL, CA, NY, GA, VA, NJ, MO, PA, WA,
Lot
Batch Number 873874875R Package including 6 needles: UDI: (01)0 7340048 30034 6 (17) 260301 (10) 873874875R (240) 911933 Package including 1 needle: UDI: (01)0 7340048 30800 7 (17) 260301 (10) 873874875R (240) 911761 Inner sterile bag including 1 needle: UDI: (01)0 7340048 30801 4 (17) 260301 (10) 873874875R (240) 912465
Risk classification
Class I
Area (FDA)
device
Country
Sweden

Related recalls

Questions about this recall

Was Elekta Disposable Biopsy Needle Kit for Leksell Stereotactic System, REF 911933, Diameter / 14 Gauge, containing 6 inner boxes labeled REF 911761, Disposable Biopsy Needle Kit for Leksell Stereotactic System. Each inner box of REF 911761 contains a Tyvek bag labeled REF 912465, Disposable Biopsy Needle and Aspirator Tube. The system is used for for localization and diagnosis of intracranial disorders and their surgical treatment including radiosurgery and stereotactic radiation therapy. recalled?

Yes. FDA published recall Z-0227-2025 on November 13, 2024. Reported hazard: Elekta has become aware that Disposable Biopsy Needles can contain some microscopic debris on the inside of the biopsy needle. The material in the debris is stainless steel, same material as the biopsy needle. No debris has been found on the outer parts of the biopsy needles.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail