Mindray A3 Anesthesia Delivery System Manufactured in China for Mindray DS USA Inc. 800 MacArthur Boulevard, Mahwah, NJ 07430 This is a device used to administer to a patient continuously or intermittently a general inhalation anesthetic and to maintain a patient's ventilation.
Published November 28, 2012 · Recall initiated: August 13, 2012
Summary
FDA records medical device Class I recall Z-0290-2013, published November 28, 2012 and initiated August 13, 2012. Recalling firm: Mindray DS USA, Inc. d.b.a. Mindray North America; Status: Terminated; Reported quantity: 69 units (61 US - 8 Foreign); Lot: 0633F-01000-00 Various serial numbers 0633F-01000-00 FA-1A000033 0633F-01000-00 FA-1A000023 0633F-01000-00 FA-21000072 0633F-01000-00 FA-1A000029 0633F-01000-00 FA-21000085 0633F-01000-00 FA-19000011 0633F-01000-00 FA-19000008 0633F-01000-00 FA-19000009 0633F-01000-00 FA-1A000027 0633F-01000-00 FA-1A000047 0633F-01000-00 FA-1A000055 0633F-01000-00 FA-1A000041 0633F-01000-00 FA-1A000053 0633F-01000-00 FA-1B000056 0633F-01000-00 FA-21000080 0633F-01000-00 FA-19000007 0633F-01000-00 FA-19000013 0633F-01000-00 FA-21000066 0633F-01000-00 FA-21000078 0633F-01000-00 FA-19000012 0633F-01000-00 FA-19000017 0633F-01000-00 FA-1A000021 0633F-01000-00 FA-1A000028 0633F-01000-00 FA-1A000038 0633F-01000-00 FA-1A000045 0633F-01000-00 FA-1A000035 0633F-01000-00 FA-1A000037 0633F-01000-00 FA-1A000026 0633F-01000-00 FA-1A000030 0633F-01000-00 FA-1A000048 0633F-01000-00 FA-1A000040 0633F-01000-00 FA-1B000059 0633F-01000-00 FA-1A000034 0633F-01000-00 FA-1A000049 0633F-01000-00 FA-21000062 0633F-01000-00 FA-21000064 0633F-01000-00 FA-21000074 0633F-01000-00 FA-21000082 0633F-01000-00 FA-21000090 0633F-01000-00 FA-1A000022 0633F-01000-00 FA-1A000051 0633F-01000-00 FA-1A000052 0633F-01000-00 FA-1A000042 0633F-01000-00 FA-1A000050 0633F-01000-00 FA-1B000057 0633F-01000-00 FA-1B000061 0633F-01000-00 FA-1A000025 0633F-01000-00 FA-1A000031 0633F-01000-00 FA-1A000036 0633F-01000-00 FA-1A000044 0633F-01000-00 FA-21000076 0633F-01000-00 FA-21000077 0633F-01000-00 FA-21000088 0633F-01000-00 FA-1A000043 0633F-01000-00 FA-1A000032 0633F-01000-00 FA-1A000046 0633F-01000-00 FA-1B000058 0633F-01000-00 FA-1B000060 0633F-01000-00 FA-19000010 0633F-01000-00 FA-19000015 0633F-01000-00 FA-21000086. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Mindray has identified the potential for a system leak resulting an issue related to the carbon dioxide absorber canister gasket on A3/A5 Anesthesia Delivery Systems. A potential system leak resulting from the canister gasket will likely present itself during the Automatic Circuit Leak and Compliance Test performed at startup and the Manual Leak Test recommended to be performed before each case.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Mindray DS USA, Inc. d.b.a. Mindray North America
- Distributor
- Worldwide distribution: USA (nationwide) including Puerto Rico and countries of: Australia, Chile, Colombia, and Venezuela.
- Lot
- 0633F-01000-00 Various serial numbers 0633F-01000-00 FA-1A000033 0633F-01000-00 FA-1A000023 0633F-01000-00 FA-21000072 0633F-01000-00 FA-1A000029 0633F-01000-00 FA-21000085 0633F-01000-00 FA-19000011 0633F-01000-00 FA-19000008 0633F-01000-00 FA-19000009 0633F-01000-00 FA-1A000027 0633F-01000-00 FA-1A000047 0633F-01000-00 FA-1A000055 0633F-01000-00 FA-1A000041 0633F-01000-00 FA-1A000053 0633F-01000-00 FA-1B000056 0633F-01000-00 FA-21000080 0633F-01000-00 FA-19000007 0633F-01000-00 FA-19000013 0633F-01000-00 FA-21000066 0633F-01000-00 FA-21000078 0633F-01000-00 FA-19000012 0633F-01000-00 FA-19000017 0633F-01000-00 FA-1A000021 0633F-01000-00 FA-1A000028 0633F-01000-00 FA-1A000038 0633F-01000-00 FA-1A000045 0633F-01000-00 FA-1A000035 0633F-01000-00 FA-1A000037 0633F-01000-00 FA-1A000026 0633F-01000-00 FA-1A000030 0633F-01000-00 FA-1A000048 0633F-01000-00 FA-1A000040 0633F-01000-00 FA-1B000059 0633F-01000-00 FA-1A000034 0633F-01000-00 FA-1A000049 0633F-01000-00 FA-21000062 0633F-01000-00 FA-21000064 0633F-01000-00 FA-21000074 0633F-01000-00 FA-21000082 0633F-01000-00 FA-21000090 0633F-01000-00 FA-1A000022 0633F-01000-00 FA-1A000051 0633F-01000-00 FA-1A000052 0633F-01000-00 FA-1A000042 0633F-01000-00 FA-1A000050 0633F-01000-00 FA-1B000057 0633F-01000-00 FA-1B000061 0633F-01000-00 FA-1A000025 0633F-01000-00 FA-1A000031 0633F-01000-00 FA-1A000036 0633F-01000-00 FA-1A000044 0633F-01000-00 FA-21000076 0633F-01000-00 FA-21000077 0633F-01000-00 FA-21000088 0633F-01000-00 FA-1A000043 0633F-01000-00 FA-1A000032 0633F-01000-00 FA-1A000046 0633F-01000-00 FA-1B000058 0633F-01000-00 FA-1B000060 0633F-01000-00 FA-19000010 0633F-01000-00 FA-19000015 0633F-01000-00 FA-21000086
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABS-200 Chemistry Analyzer Designed for clinical laboratory use.April 3, 2013
- FDAMindray DS USA, Inc., DPM Central Monitoring System with ambulatory Telepack. Part numbers: 300BF-PA00002, andFebruary 20, 2013
- FDAMindray DPM5 Monitor, a vital signs monitor used on human patients in healthcare facilities.February 6, 2013
- FDADPM Central Station Mindray DS USA, Inc. 800 MacArthur Boulevard, Mahwah, NJ 07430 The Central MonitoringJanuary 9, 2013
- FDAGE Healthcare Brivo XR 285 Mobile general purpose radiographic imaging of the human head and body.November 28, 2012
- FDAMindray A5 Anesthesia Delivery System Manufactured in China for Mindray DS USA Inc. 800 MacArthur Boulevard,November 28, 2012
Questions about this recall
Was Mindray A3 Anesthesia Delivery System Manufactured in China for Mindray DS USA Inc. 800 MacArthur Boulevard, Mahwah, NJ 07430 This is a device used to administer to a patient continuously or intermittently a general inhalation anesthetic and to maintain a patient's ventilation. recalled?
Yes. FDA published recall Z-0290-2013 on November 28, 2012. Reported hazard: Mindray has identified the potential for a system leak resulting an issue related to the carbon dioxide absorber canister gasket on A3/A5 Anesthesia Delivery Systems. A potential system leak resulting from the canister gasket will likely present itself during the Automatic Circuit Leak and Compliance Test performed at startup and the Manual Leak Test recommended to be performed before each case.
What should I do if I have this product?
Voluntary: Firm initiated · Letter