GE Healthcare Brivo XR 285 Mobile general purpose radiographic imaging of the human head and body.

Published November 28, 2012 · Recall initiated: September 20, 2012

Summary

FDA records medical device Class II recall Z-0200-2013, published November 28, 2012 and initiated September 20, 2012. Recalling firm: GE Healthcare, LLC; Status: Terminated; Reported quantity: 364 total units installed in US; Lot: part numbers:5555000, 5555000-2. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

It was discovered GE Healthcare has become aware of a firmware issue on the mobile radiographic product, Brivo XR285amx, Optima XR200amx, Optima XR22Oamx and Optima XR200amx with Digital Upgrade, related to Technique Accuracy as mandated by the US Code of Federal Regulations (21CFR) and IEC 60601-2-54.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
GE Healthcare, LLC
Distributor
Nationwide Distribution
Lot
part numbers:5555000, 5555000-2
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was GE Healthcare Brivo XR 285 Mobile general purpose radiographic imaging of the human head and body. recalled?

Yes. FDA published recall Z-0200-2013 on November 28, 2012. Reported hazard: It was discovered GE Healthcare has become aware of a firmware issue on the mobile radiographic product, Brivo XR285amx, Optima XR200amx, Optima XR22Oamx and Optima XR200amx with Digital Upgrade, related to Technique Accuracy as mandated by the US Code of Federal Regulations (21CFR) and IEC 60601-2-54.

What should I do if I have this product?

Voluntary: Firm initiated · Letter