Trilogy Evo O2. Model Numbers: FP2100X10, DS2100X11B, IN2100X15B, DE2100X13B, IA2100X15B, EU2100X15B, LA2100X15B, EE2100X15B, FX2100X15B, ES2100X15B, ND2100X15B, FR2100X14B, IT2100X21B, JP2100X16B, IN2100X19, EU2100X19. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.
Published November 20, 2024 · Recall initiated: October 7, 2024
Summary
FDA records medical device Class I recall Z-0290-2025, published November 20, 2024 and initiated October 7, 2024. Recalling firm: Philips Respironics, Inc.; Status: Ongoing; Reported quantity: 24,249 units; Lot: Model Number (UDI-DI Number): DE2100X13B (606959054103); DS2100X11B (606959051997); EE2100X15B (606959055698); ES2100X15B (606959054097); EU2100X15B (606959055674); EU2100X19 (606959055681); FP2100X10 (606959062085); FR2100X14B (606959055643); FX2100X15B (606959061019); IA2100X15B (606959055636); IN2100X15B (606959054059); IN2100X19 (606959055575); IT2100X21B (606959055438); JP2100X16B (606959055582); LA2100X15B (606959055599); ND2100X15B (606959055650). All Serial Numbers. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Respironics, Inc.
- Distributor
- Worldwide distribution - US Nationwide and the countries of AE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JE, JO, JP, KE, KR, KW, LB, LT, LY, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PT, PY, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, TZ, UA, UY, VN, YT, ZA, ZW.
- Lot
- Model Number (UDI-DI Number): DE2100X13B (606959054103); DS2100X11B (606959051997); EE2100X15B (606959055698); ES2100X15B (606959054097); EU2100X15B (606959055674); EU2100X19 (606959055681); FP2100X10 (606959062085); FR2100X14B (606959055643); FX2100X15B (606959061019); IA2100X15B (606959055636); IN2100X15B (606959054059); IN2100X19 (606959055575); IT2100X21B (606959055438); JP2100X16B (606959055582); LA2100X15B (606959055599); ND2100X15B (606959055650). All Serial Numbers.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPhilips Respironics Trilogy EV300, Software Version 1.05.15.00. Continuous home-use ventilator device.April 1, 2026
- FDAPhilips Respironics Trilogy Evo Universal, Software Version 1.05.15.00. Continuous home-use ventilator device.April 1, 2026
- FDABiPAP A40 Ventilator. Used to provide noninvasive ventilation support for the purpose of treating adult andJuly 9, 2025
- FDAGarbin Evo. Model Number: LD2110X23B. Intended for invasive and non-invasive ventilation of pediatric throughNovember 20, 2024
- FDA154784 1 / 5LB IMPERIALFRESH PROC FAJITA GRN PEPRED PEP W All lot codesNovember 20, 2024
- FDATrilogy Evo. Model Numbers: DS2110X11B, UDS2110X11B, RDS2110X11B, CA2110X12B, JP2110X16B, IN2110X15B,November 20, 2024
Questions about this recall
Was Trilogy Evo O2. Model Numbers: FP2100X10, DS2100X11B, IN2100X15B, DE2100X13B, IA2100X15B, EU2100X15B, LA2100X15B, EE2100X15B, FX2100X15B, ES2100X15B, ND2100X15B, FR2100X14B, IT2100X21B, JP2100X16B, IN2100X19, EU2100X19. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings. recalled?
Yes. FDA published recall Z-0290-2025 on November 20, 2024. Reported hazard: Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.
What should I do if I have this product?
Voluntary: Firm initiated · Letter