Back to search
FDAFDA · Food and drugs·Recall number Z-0295-2025

Garbin Evo. Model Number: LD2110X23B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.

Published November 20, 2024 · Recall initiated: October 7, 2024

Summary

FDA records medical device Class I recall Z-0295-2025, published November 20, 2024 and initiated October 7, 2024. Recalling firm: Philips Respironics, Inc.; Status: Ongoing; Reported quantity: 1,322 units; Lot: Model No. LD2110X23B; UDI-DI: 606959057432; All Serial Numbers. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips Respironics, Inc.
Distributor
Worldwide distribution - US Nationwide and the countries of AE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JE, JO, JP, KE, KR, KW, LB, LT, LY, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PT, PY, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, TZ, UA, UY, VN, YT, ZA, ZW.
Lot
Model No. LD2110X23B; UDI-DI: 606959057432; All Serial Numbers.
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Garbin Evo. Model Number: LD2110X23B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings. recalled?

Yes. FDA published recall Z-0295-2025 on November 20, 2024. Reported hazard: Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.

What should I do if I have this product?

Voluntary: Firm initiated · Letter