Fresenius 132 Gallon Dry Acid Dissolution Unit, P/N 160147. For Use in the Preparation of acid concentrate for hemodialysis.
Published November 25, 2015 · Recall initiated: October 23, 2015
Summary
FDA records medical device Class II recall Z-0295-2016, published November 25, 2015 and initiated October 23, 2015. Recalling firm: Fresenius Medical Care Renal Therapies Group, LLC; Status: Terminated; Reported quantity: 348 units; Lot: All 132 Gallon Granuflo Dissolution Units and 132 Gallon Dry Acid Dissolution Units with the H218 Motor and Modified H218 Mixer Motors. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Dissolution units and replacement motors units do not provide full thermal protection for both the Main and Starter coils resulting in the generation of smoke and/or potentially lead to fire.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Fresenius Medical Care Renal Therapies Group, LLC
- Distributor
- Nationwide Distribution.
- Lot
- All 132 Gallon Granuflo Dissolution Units and 132 Gallon Dry Acid Dissolution Units with the H218 Motor and Modified H218 Mixer Motors.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADELFLEX Peritoneal Dialysis Solution with 1.5% Dextrose, Low Magnesium/ Low Calcium and Attached stay safeNovember 11, 2026
- FDADELFLEX Peritoneal Dialysis Solution with 1.5% Dextrose, Low Magnesium/ Low Calcium and Attached stay safeNovember 11, 2026
- FDADELFLEX Peritoneal Dialysis Solution with 2.5% Dextrose, Low Magnesium/ Low Calcium and Attached stay safeNovember 11, 2026
- FDADELFLEX Peritoneal Dialysis Solution with 2.5% Dextrose, Low Magnesium/ Low Calcium and Attached stay"safeNovember 11, 2026
- FDACIVCO Sales Demo Kit. REF DEMOGPGUIDE Large Gauge Infiniti Plus Needle Guides, used in needle-guided (orNovember 25, 2015
- FDAFresenius Modified Mixer Motor (for 132 Gallon Dissolution Units), P/N 160075. For Use in the Preparation ofNovember 25, 2015
Questions about this recall
Was Fresenius 132 Gallon Dry Acid Dissolution Unit, P/N 160147. For Use in the Preparation of acid concentrate for hemodialysis. recalled?
Yes. FDA published recall Z-0295-2016 on November 25, 2015. Reported hazard: Dissolution units and replacement motors units do not provide full thermal protection for both the Main and Starter coils resulting in the generation of smoke and/or potentially lead to fire.
What should I do if I have this product?
Voluntary: Firm initiated · Letter