FDAFDA · Food and drugs·Recall number Z-0296-2016
Fresenius Modified Mixer Motor (for 132 Gallon Dissolution Units), P/N 160075. For Use in the Preparation of acid concentrate for hemodialysis.
Published November 25, 2015 · Recall initiated: October 23, 2015
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Dissolution units and replacement motors units do not provide full thermal protection for both the Main and Starter coils resulting in the generation of smoke and/or potentially lead to fire.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Fresenius Medical Care Renal Therapies Group, LLC
- Distributor
- Nationwide Distribution.
- Lot
- All 132 Gallon Granuflo Dissolution Units and 132 Gallon Dry Acid Dissolution Units with the H218 Motor and Modified H218 Mixer Motors.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States