Back to search
FDAFDA · Food and drugs·Recall number Z-0315-2025

intelliPATH Universal HRP Detection Kit, REF: IPK5011G80, containing DAB Buffer, Catalog Number: IPBF5009G20; ONCORE Pro DAB Chromogen, REF: OPRI6056KT180, containing DAB Buffer, Catalog Number: OPRI6055T60; IP DAB Chromogen Kit 80 ml, REF: IPK5010G80, containing DAB Buffer, Catalog Number: IPBF5009G20; DAB kit, REF: OPRI6056KT180, containing DAB Buffer, Catalog Number: OPRI6055T60 Folate Receptor alpha IHC Assay Kit, REF: IPI4006KG10, containing DAB Buffer, Catalog Number: IPBF5009G20

Published November 13, 2024 · Recall initiated: September 25, 2024

Summary

FDA records medical device Class II recall Z-0315-2025, published November 13, 2024 and initiated September 25, 2024. Recalling firm: Biocare Medical, LLC; Status: Ongoing; Reported quantity: 976; Lot: REF/UDI-DI/Kit Lot(DAB Lot): IPK5011G80/00847627009005/042324(012424A),060523(051823A), 082923(062923A), 011624(072723A), 042324(012424A), 080524(032624A); OPRI6056KT180/00847627024985/020624(111623A),032224(111623A-3); IPK5010G80/00847627008992/071923(051823A), 091923(081023A), 112823(101923A), 100424(011524A), 041624(022624A); OPRI6056KT180/00847627024985/071924(061824A); IPI4006KG10/00847627008985/080323(062923A), 102723(081023A). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Kits containing the DAB chromogen buffer, intended for use in either manual or automated Immunohistochemistry (IHC) staining protocols, may produce weak to no staining, which may lead to test failure requiring the test to be repeated.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Biocare Medical, LLC
Distributor
US Nationwide distribution in the states of MA, AZ, CT, NY, FL, NC, IN, IL, MO, CO, TX, CA, MI, PR, PA, MD, AL, OH, KY.
Lot
REF/UDI-DI/Kit Lot(DAB Lot): IPK5011G80/00847627009005/042324(012424A),060523(051823A), 082923(062923A), 011624(072723A), 042324(012424A), 080524(032624A); OPRI6056KT180/00847627024985/020624(111623A),032224(111623A-3); IPK5010G80/00847627008992/071923(051823A), 091923(081023A), 112823(101923A), 100424(011524A), 041624(022624A); OPRI6056KT180/00847627024985/071924(061824A); IPI4006KG10/00847627008985/080323(062923A), 102723(081023A)
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was intelliPATH Universal HRP Detection Kit, REF: IPK5011G80, containing DAB Buffer, Catalog Number: IPBF5009G20; ONCORE Pro DAB Chromogen, REF: OPRI6056KT180, containing DAB Buffer, Catalog Number: OPRI6055T60; IP DAB Chromogen Kit 80 ml, REF: IPK5010G80, containing DAB Buffer, Catalog Number: IPBF5009G20; DAB kit, REF: OPRI6056KT180, containing DAB Buffer, Catalog Number: OPRI6055T60 Folate Receptor alpha IHC Assay Kit, REF: IPI4006KG10, containing DAB Buffer, Catalog Number: IPBF5009G20 recalled?

Yes. FDA published recall Z-0315-2025 on November 13, 2024. Reported hazard: Kits containing the DAB chromogen buffer, intended for use in either manual or automated Immunohistochemistry (IHC) staining protocols, may produce weak to no staining, which may lead to test failure requiring the test to be repeated.

What should I do if I have this product?

Voluntary: Firm initiated · Letter