FDAFDA · Food and drugs·Recall number Z-0317-2025
The FLUOROspeed is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. FLUOROspeed may be used for emergency treatment on an outpatient basis, as well as for bedside examinations.
Published November 13, 2024 · Recall initiated: September 6, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
X-ray R/F system software with certain versions of the control board may occasionally fail to display images when switching from Fluoroscopy mode to Radiography mode, which may lead to the possibility of delay or cancellation of the examination.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION
- Distributor
- US Nationwide distribution in the states of CA, CT, MA, WV, OH, NY, IA, NE, MN, IL, FL, NC, HI, LA, MD, MT, WA, AR, TX, TN, CO.
- Lot
- UDI-DI: 04540217065271. Serial Numbers: MQ493E3E3001, MQ493E3E3002, MQ493E3D6002, MQ493E3D5001, MQ493E3D8002, MQ493E3E1002, MQ493E3D4002, MQ493E3D6003, MQ493E3DA002, MQ493E3E2001, MQ493E3E2002, MQ493E3E1001, MQ493E3DC001, MQ493E3D9001, MQ493E3D4005, MQ493E3D7001, MQ493E3DB002, MQ493E3D2002, MQ493E3D4001, MQ493E3D7002, MQ493E3D4003, MQ493E3D4004, MQ493E3D2001, MQ493E3D8001, MQ493E3D5002, MQ493E3DB001, MQ493E3D2003, MQ493E3DA001, MQ493E3D6001, MQ493E3E1003
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Japan