Back to search
FDAFDA · Food and drugs·Recall number Z-0333-2024

Gastroscope-endoscopic ultrasound (EUS) guided fine needle aspiration (FNA) and for endoscopic surgery within the upper digestive tract. Model Numbers: GF-UC140P-AL5, GF-UCT180, GF-UE160-AL5.

Published November 29, 2023 · Recall initiated: September 25, 2023

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Aizu Olympus Co., Ltd.
Distributor
Worldwide - US Nationwide distribution.
Lot
Model Numbers: GF-UC140P-AL5 (04953170356322), GF-UCT180 (04953170356339), GF-UE160-AL5 (04953170356261).
Risk classification
Class II
Area (FDA)
device
Country
Japan