Back to search
FDAFDA · Food and drugs·Recall number Z-0342-2024

Signmoidovideoscope-For endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum and sigmoid colon). Model Number: PCF-S.

Published November 29, 2023 · Recall initiated: September 25, 2023

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Aizu Olympus Co., Ltd.
Distributor
Worldwide - US Nationwide distribution.
Lot
Model Number (UDI-DI): PCF-S (04953170339912).
Risk classification
Class II
Area (FDA)
device
Country
Japan