Guider Softip XF; GUIDER/MP XF/8FR/100 cm, Model number: M003101480. Cardiology: Boston Scientific Guider Softip XF Guide Catheters are intended to facilitate the placement of interventional devices into the neurovascular system.

Published November 9, 2016 · Recall initiated: October 5, 2016

Summary

FDA records medical device Class II recall Z-0334-2017, published November 9, 2016 and initiated October 5, 2016. Recalling firm: Stryker Neurovascular; Status: Terminated; Reported quantity: 1478 devices total, all model numbers and lots.; Lot: Lot 18713574. Expiration 11/30/18. Lot s18849984, 18848795. Expiration 01/31/19. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Firm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Stryker Neurovascular
Distributor
Worldwide Distribution - US including AK, IL, OH and Internationally to: ARGENTINA AUSTRALIA AUSTRIA BELGIUM BRAZIL CANADA CHILE CZECH REPUBLIC ECUADOR EGYPT FRANCE GERMANY GREECE HUNGARY INDIA IRAN ISRAEL ITALIA JAPAN MEXICO NETHERLANDS NORWAY PERU PHILIPPINES POLAND RUSSIA SERBIA SINGAPORE SLOVAKIA SOUTH AFRICA SOUTH KOREA SPAIN SWEDEN SWITZERLAND THAILAND TUNISIA TURKEY UK URUGUAY VIETNAM.
Lot
Lot 18713574. Expiration 11/30/18. Lot s18849984, 18848795. Expiration 01/31/19.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Guider Softip XF; GUIDER/MP XF/8FR/100 cm, Model number: M003101480. Cardiology: Boston Scientific Guider Softip XF Guide Catheters are intended to facilitate the placement of interventional devices into the neurovascular system. recalled?

Yes. FDA published recall Z-0334-2017 on November 9, 2016. Reported hazard: Firm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product.

What should I do if I have this product?

Voluntary: Firm initiated · Letter