The Synchro SELECT Guidewires are a steerable guidewire family with a shapeable tip and are available in straight and pre-shaped versions. The outside diameters of the guidewires are 0.014in. The guidewires are compatible with existing microcatheters used in common procedures such as those used in endovascular diagnosis and therapy of neurovascular disease. The distal portion of the guidewire tip is radiopaque. For lubricity, the distal portion of the device is coated with a hydrophilic polymer and the proximal portion of the guidewire is coated with polytetrafluoroethylene (PTFE). The torque device included with the guidewire attaches to the proximal end of the wire and functions as a steering guide. Rotation of this device facilitates guidewire placement into the appropriate vessel by precise directional manipulation of the guidewire tip. The introducer included with the guidewire is intended to aid insertion of the guidewire into the catheter hub and/or hemostasis valve.
Published June 12, 2024 · Recall initiated: April 18, 2024
Summary
FDA records medical device Class II recall Z-1940-2024, published June 12, 2024 and initiated April 18, 2024. Recalling firm: Stryker Neurovascular; Status: Ongoing; Reported quantity: 41,074; Lot: Catalog/UDI-DI/Lots: SSFT215PRE/07613327508598/0000139994, 0000144321, 0000146842, 0000154408; SSFT215STR/07613327508574/0000122897, 0000123151, 0000123405, 0000124931, 0000125040, 0000125719, 0000125720, 0000125787, 0000126356, 0000129220, 0000129221, 0000130861, 0000131723, 0000132954, 0000133179, 0000137045, 0000138580, 0000138581, 0000142280, 0000142281, 0000143349, 0000143584; SSFT300STR/07613327508581/0000144375, 0000155022; SSTD215PRE/07613327508512/0000122145, 0000122866, 0000127369, 0000139986, 0000140672, 0000141825, 0000141826, 0000143802, 0000146926, 0000149561, 0000154409, 0000154410, 0000154411, 0000176472; SSTD215STR/07613327508499/0000122481, 0000122485, 0000122650, 0000123069, 0000123070, 0000123143, 0000125409, 0000127494, 0000127495, 0000128254, 0000128446, 0000128447, 0000131661, 0000131667, 0000131668, 0000133396, 0000133563, 0000133564, 0000140971, 0000141392, 0000141393, 0000142532, 0000143125, 0000143209, 0000143806, 0000143834, 0000144140, 0000148427, 0000148428, 0000148477, 0000149560, 0000154679, 0000154680, 0000154681, 0000155021, 0000155662, 0000156792, 0000164588, 0000164589, 0000164591, 0000172898, 0000172899, 0000172900, 0000172911, 0000175987, 0000176018, 0000176019; SSTD300STR/07613327508505/0000145095, 0000145694, 0000151281; SSUP215PRE/07613327508635/0000144317, 0000148425; SSUP300PRE/07613327508642/0000145692, 0000151280; SSUP215STR/07613327508611/0000128253, 0000129625, 0000137050, 0000140671, 0000141784, 0000143342, 0000144318, 0000145693, 0000145875, 0000164662, 0000165477; SSUP300STR/07613327508628/0000121919, 0000141789, 0000145691. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro Guidewire that contain an older version of the introducer accessory. Due to variation in the manufacturing process of the supplier of the introducer accessory, certain lots of introducers have sharper than intended edges that can peel off the PTFE coating when physicians use a technique known as backloading. Users with impacted product in their inventory are cautioned not to use this backloading technique.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Neurovascular
- Distributor
- Worldwide - US Nationwide distribution including in the states of TX, NY, MD, CA, KY, WI, OR, NE, SC, MA, MO, CT, ID, PA, TN, FL, NC, AL, IA, OK, VA, IL, OH, NH, MN, ND, KS, NJ, WV, IN, GA, HI, AR, UT, LA, MI, WA, NV, CO, AZ, DC, VT, ME, DE, AK, WY, MT, NM, SD, MS, RI, GU and the countries of ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BRAZIL, BULGARIA, CANADA, CHINA, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, ICELAND, INDIA, IRAN, IRELAND, ISRAEL, ITALIA, JAPAN, JORDAN, KUWAIT, LITHUANIA, LUXEMBOURG, MEXICO, NETHERLANDS, NORTHERN IRELAND, NORWAY, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UAE, UK.
- Lot
- Catalog/UDI-DI/Lots: SSFT215PRE/07613327508598/0000139994, 0000144321, 0000146842, 0000154408; SSFT215STR/07613327508574/0000122897, 0000123151, 0000123405, 0000124931, 0000125040, 0000125719, 0000125720, 0000125787, 0000126356, 0000129220, 0000129221, 0000130861, 0000131723, 0000132954, 0000133179, 0000137045, 0000138580, 0000138581, 0000142280, 0000142281, 0000143349, 0000143584; SSFT300STR/07613327508581/0000144375, 0000155022; SSTD215PRE/07613327508512/0000122145, 0000122866, 0000127369, 0000139986, 0000140672, 0000141825, 0000141826, 0000143802, 0000146926, 0000149561, 0000154409, 0000154410, 0000154411, 0000176472; SSTD215STR/07613327508499/0000122481, 0000122485, 0000122650, 0000123069, 0000123070, 0000123143, 0000125409, 0000127494, 0000127495, 0000128254, 0000128446, 0000128447, 0000131661, 0000131667, 0000131668, 0000133396, 0000133563, 0000133564, 0000140971, 0000141392, 0000141393, 0000142532, 0000143125, 0000143209, 0000143806, 0000143834, 0000144140, 0000148427, 0000148428, 0000148477, 0000149560, 0000154679, 0000154680, 0000154681, 0000155021, 0000155662, 0000156792, 0000164588, 0000164589, 0000164591, 0000172898, 0000172899, 0000172900, 0000172911, 0000175987, 0000176018, 0000176019; SSTD300STR/07613327508505/0000145095, 0000145694, 0000151281; SSUP215PRE/07613327508635/0000144317, 0000148425; SSUP300PRE/07613327508642/0000145692, 0000151280; SSUP215STR/07613327508611/0000128253, 0000129625, 0000137050, 0000140671, 0000141784, 0000143342, 0000144318, 0000145693, 0000145875, 0000164662, 0000165477; SSUP300STR/07613327508628/0000121919, 0000141789, 0000145691
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAINZONE DETACHMENT SYSTEM, REF: M00345100950July 1, 2026
- FDAThe Trevo Trak 21 Microcatheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaqueOctober 16, 2024
- FDAThe Synchro2 Guidewires are a steerable guidewire family with a shapeable tip and are available in straightJune 12, 2024
- FDAThe Synchro Neuro Guidewires are a steerable guidewire family with a shapeable tip. The outside diameter ofJune 12, 2024
- FDAMEDLINE LUER SLIP DISPOSABLE SYRINGE WITHOUT NEEDLE, REF SYR101020June 12, 2024
- FDAWAVi Research EEG Desktop System is comprised of the WAVi Research EEG Desktop Software with InstructionJune 12, 2024
Questions about this recall
Was The Synchro SELECT Guidewires are a steerable guidewire family with a shapeable tip and are available in straight and pre-shaped versions. The outside diameters of the guidewires are 0.014in. The guidewires are compatible with existing microcatheters used in common procedures such as those used in endovascular diagnosis and therapy of neurovascular disease. The distal portion of the guidewire tip is radiopaque. For lubricity, the distal portion of the device is coated with a hydrophilic polymer and the proximal portion of the guidewire is coated with polytetrafluoroethylene (PTFE). The torque device included with the guidewire attaches to the proximal end of the wire and functions as a steering guide. Rotation of this device facilitates guidewire placement into the appropriate vessel by precise directional manipulation of the guidewire tip. The introducer included with the guidewire is intended to aid insertion of the guidewire into the catheter hub and/or hemostasis valve. recalled?
Yes. FDA published recall Z-1940-2024 on June 12, 2024. Reported hazard: Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro Guidewire that contain an older version of the introducer accessory. Due to variation in the manufacturing process of the supplier of the introducer accessory, certain lots of introducers have sharper than intended edges that can peel off the PTFE coating when physicians use a technique known as backloading. Users with impacted product in their inventory are cautioned not to use this backloading technique.
What should I do if I have this product?
Voluntary: Firm initiated · Letter