Epsilon Skin Tensioner:Rx, Sterile Sterilized using Gamma. Turns Red if sterilized using Gamma; Product Usage: Skin Tensioner is applied to patients scalp during follicular harvesting during hair restoration procedure. Product applies tension to scalp to facilitate follicular extractions.
Published November 27, 2013 · Recall initiated: October 13, 2013
Summary
FDA records medical device Class II recall Z-0353-2014, published November 27, 2013 and initiated October 13, 2013. Recalling firm: Restoration Robotics; Status: Terminated; Reported quantity: 515 in US. 315 International.(included all packaging configurations); Lot: Model number FP-45384. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Reports of discoloration (residue) on the pins of the skin tensioner have been received by several customers. Discoloration due to insufficient cleaning of stainless steel pins during passivation process. The residue has been identified as a compound of iron and chlorine.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Restoration Robotics
- Distributor
- Worldwide Distribution - USA Nationwide in the countries of: Korea, Japan, Singapore, Thailand, United Kingdom, Canada, and Taiwan.
- Lot
- Model number FP-45384
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAARTAS Disposable Kit ( Contains Skin Tensioner FP-45384; Inner Needle with Vacuum Tube FP-45385; DissectionNovember 27, 2013
- FDAARTAS Disposable and Reuseable Kit ( Contains Skin Tensioner FP-45384; Inner Needle with Vacuum Tube FP-45385;November 27, 2013
- FDASterile Disposable Kit ( Contains Skin Tensioner ASY-25067; Three Mark Needle Assembly ASY-25094; Nose ConeNovember 27, 2013
- FDACAS I/II Absorbers that are parts of the BleaseSirius Anesthesia Workstation, the BleaseFocus AnesthesiaNovember 27, 2013
- FDASpacelabs Healthcare qube Compact Patient Monitor, Model 91390. It is a patient monitor with arrhythmiaNovember 27, 2013
- FDAThe GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, inNovember 27, 2013
Questions about this recall
Was Epsilon Skin Tensioner:Rx, Sterile Sterilized using Gamma. Turns Red if sterilized using Gamma; Product Usage: Skin Tensioner is applied to patients scalp during follicular harvesting during hair restoration procedure. Product applies tension to scalp to facilitate follicular extractions. recalled?
Yes. FDA published recall Z-0353-2014 on November 27, 2013. Reported hazard: Reports of discoloration (residue) on the pins of the skin tensioner have been received by several customers. Discoloration due to insufficient cleaning of stainless steel pins during passivation process. The residue has been identified as a compound of iron and chlorine.
What should I do if I have this product?
Voluntary: Firm initiated · Letter