Sterile Disposable Kit ( Contains Skin Tensioner ASY-25067; Three Mark Needle Assembly ASY-25094; Nose Cone Assembly ASY-25098; Dissection 4 mm PRT-25117, follicle Trap PRT 23999): Product Usage: Skin Tensioner is applied to patients scalp during follicular harvesting during hair restoration procedure. Product applies tension to scalp to facilitate follicular extractions.

Published November 27, 2013 · Recall initiated: October 13, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Reports of discoloration (residue) on the pins of the skin tensioner have been received by several customers. Discoloration due to insufficient cleaning of stainless steel pins during passivation process. The residue has been identified as a compound of iron and chlorine.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Restoration Robotics
Distributor
Worldwide Distribution - USA Nationwide in the countries of: Korea, Japan, Singapore, Thailand, United Kingdom, Canada, and Taiwan.
Lot
Model number FP-45547
Risk classification
Class II
Area (FDA)
device
Country
United States