Yeastone Broth, 11ML, 10/BOX, used in antibiotic susceptibility testing of mycological clinical specimens
Published November 11, 2026 · Recall initiated: October 19, 2026
Summary
FDA records medical device Class II recall Z-0355-2027, published November 11, 2026 and initiated October 19, 2026. Recalling firm: Remel, Inc; Status: Ongoing; Reported quantity: 33 US, 54 OUS; Lot: YY3462 lot 440400 UDI-DI 848838018817. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Affected lot did not meet specifications. Product may report false susceptibility to antifungal agents, specifically within the Azole family.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Remel, Inc
- Distributor
- Worldwide - US Nationwide distribution in the states of MA and CA. The countries of India, Saudi Arabia, UK.
- Lot
- YY3462 lot 440400 UDI-DI 848838018817
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column, Model/Catalog Number: 1854087November 11, 2026
- FDAUnity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. Part Number: 1232040.November 11, 2026
- FDAGC Agar Base (150mm) 10/PK R04030; GC Agar Base (100mm) 10/PK R01460July 8, 2026
- FDAYeastone Broth, 11ML, 10/BOX YY3462April 22, 2026
- FDAThermo SCIENTIFIC Sensititre Cation Adjusted AutoRead Mueller-Hinton Broth w/ TES w/ Lysed Horse Blood, REFJune 18, 2025
- FDAZebra Thermal Barcode Printer GX430t, Model no. YSW 1000 or YSW 2000, may have been sold as an optionalMarch 26, 2025
Questions about this recall
Was Yeastone Broth, 11ML, 10/BOX, used in antibiotic susceptibility testing of mycological clinical specimens recalled?
Yes. FDA published recall Z-0355-2027 on November 11, 2026. Reported hazard: Affected lot did not meet specifications. Product may report false susceptibility to antifungal agents, specifically within the Azole family.
What should I do if I have this product?
Voluntary: Firm initiated · Letter