Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. Part Number: 1232040.
Published November 11, 2026 · Recall initiated: October 2, 2026
Summary
FDA records medical device Class II recall Z-0356-2027, published November 11, 2026 and initiated October 2, 2026. Recalling firm: CORIN MEDICAL, LTD.; Status: Ongoing; Reported quantity: 22 units; Lot: Part Number: 1232040. UDI-DI/GTIN; 05056139239656. Lot Code: 097921. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential malfunction of the Modular Handles where the locking mechanism may fail. Any mating instruments already engaged to the handle will not be able to be disengaged, and the handle will be unable to fitted securely to any unengaged mating instruments.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- CORIN MEDICAL, LTD.
- Distributor
- Worldwide - US Nationwide distribution in the states of CA, GA, IL, MI, NC, RI, TX, UT and the countries of Belgium, France, Germany, Hungary.
- Lot
- Part Number: 1232040. UDI-DI/GTIN; 05056139239656. Lot Code: 097921.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United Kingdom
Related recalls
- FDAYeastone Broth, 11ML, 10/BOX, used in antibiotic susceptibility testing of mycological clinical specimensNovember 11, 2026
- FDAintelerad IntelePACS softwareNovember 11, 2026
- FDATruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column, Model/Catalog Number: 1854087November 11, 2026
- FDAHalyard ARTHROSCOPY PACK. Model/Catalog Number: SAMM014-21.November 11, 2026
- FDAOperating Table TruSystem 7000 U (dV), Model/Catalog Number: 1723633November 11, 2026
- FDALIAISON PLEX Gram-Negative Blood Culture Assay Catalog No. 20-005-303November 11, 2026
Questions about this recall
Was Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. Part Number: 1232040. recalled?
Yes. FDA published recall Z-0356-2027 on November 11, 2026. Reported hazard: Potential malfunction of the Modular Handles where the locking mechanism may fail. Any mating instruments already engaged to the handle will not be able to be disengaged, and the handle will be unable to fitted securely to any unengaged mating instruments.
What should I do if I have this product?
Voluntary: Firm initiated · Letter