EPIQ 7 Ultrasound System, Model 795200, 795201, 795202. Intended for diagnostic ultrasound imaging and fluid flow analysis.
Published November 27, 2013 · Recall initiated: October 17, 2013
Summary
FDA records medical device Class II recall Z-0368-2014, published November 27, 2013 and initiated October 17, 2013. Recalling firm: Philips Medical Systems North America Inc.; Status: Terminated; Reported quantity: 306 units total (206 international; 100 in the US).; Lot: TRACE ITEM SERIAL NUMBERS in the UNITED STATES: US713B0171, US713B0207, US813B0017, US813B0018, US813B0019, US813B0020, US813B0022, US813B0023, US813B0024, US813B0025, US813B0122, US813B0123, US813B0125, US813B0127, US813B0128, US813B0129, US813B0131, US813B0132, US813B0134, US813B0135, US813B0233, US813B0234, US813B0235, US813B0236, US813B0237, US813B0238, US813B0239, US813B0240, US813B0423, US813B0424, US813B0425, US813B0426, US813B0427, US813B0428, US813B0429, US813B0430, US813B0431, US813B0432, US813B0433, US813B0434, US913B0033, US913B0034, US913B0035, US913B0036, US913B0037, US913B0038, US913B0039, US913B0040, US913B0057, US913B0060, US913B0061, US913B0062, US913B0063, US913B0064, US913B0065, US913B0066, US913B0106, US913B0107, US913B0108, US913B0109, US913B0110, US913B0111, US913B0112, US913B0113, US913B0144, US913B0176, US913B0201, US913B0256, US913B0257, US913B0350, US913B0363, US913B0364, US913B0365, US913B0366, US913B0367, US913B0368, US913B0369, US913B0380, US913B0384, US913B0395, US913B0396, US913B0397, US913B0398, US913B0399, US913B0400, US913B0401, US913B0402, US913B0409, US913B0441, US913B0444, US913B0445, US913B0446, US913B0447, US913B0503, US913B0522, US913B0614, US913B0615, US913B0616, US913B0618, US913B0619, and USO13B0043. TRACE ITEM SERIAL NUMBERS INTERNATIONAL: US513B0224, US513B0295, US513B0387, US513B0560, US613B0299, US713B0085, US713B0086, US713B0118, US713B0119, US713B0120, US713B0122, US713B0123, US713B0126, US713B0149, US713B0171, US713B0172, US713B0192, US713B0193, US713B0194, US713B0195, US713B0204, US713B0205, US713B0206, US713B0208, US713B0324, US713B0325, US713B0358, US713B0359, US713B0437, US713B0439, US713B0440, US713B0532, US713B0537, US713B0573, US713B0574, US713B0575, US713B0576, US713B0577, US713B0578, US713B0579, US713B0580, US713B0581, US713B0582, US713B0583, US713B0584, US713B0586, US713B0587, US713B0588, US713B0589, US713B0590, US713B0592, US713B0593, US713B0594, US713B0595, US813B0015, US813B0016, US813B0032, US813B0037, US813B0038, US813B0039, US813B0121, US813B0151, US813B0152, US813B0153, US813B0154, US813B0155, US813B0156, US813B0157, US813B0158, US813B0159, US813B0160, US813B0161, US813B0162, US813B0163, US813B0164, US813B0183, US813B0184, US813B0185, US813B0186, US813B0187, US813B0188, US813B0189, US813B0190, US813B0191, US813B0192, US813B0193, US813B0194, US813B0195, US813B0196, US813B0232, US813B0396, US813B0397, US813B0398, US813B0399, US813B0400, US813B0401, US813B0402, US813B0403, US813B0404, US813B0405, US813B0406, US813B0407, US813B0408, US813B0409, US813B0410, US813B0411, US813B0468, US813B0469, US813B0470, US813B0471, US813B0472, US813B0473, US813B0474, US813B0475, US813B0476, US813B0477, US813B0478, US813B0479, US813B0480, US913B0026, US913B0027, US913B0028, US913B0029, US913B0030, US913B0031, US913B0032, US913B0041, US913B0051, US913B0059, US913B0085, US913B0086, US913B0101, US913B0102, US913B0103, US913B0104, US913B0105, US913B0130, US913B0132, US913B0133, US913B0134, US913B0135, US913B0137, US913B0138, US913B0139, US913B0140, US913B0141, US913B0142, US913B0143, US913B0177, US913B0178, US913B0179, US913B0185, US913B0202, US913B0216, US913B0217, US913B0218, US913B0219, US913B0220, US913B0221, US913B0222, US913B0223, US913B0232, US913B0252, US913B0253, US913B0254, US913B0255, US913B0259, US913B0260, US913B0299, US913B0300, US913B0301, US913B0302, US913B0303, US913B0316, US913B0317, US913B0318, US913B0319, US913B0332, US913B0333, US913B0336, US913B0337, US913B0338, US913B0339, US913B0340, US913B0349, US913B0359, US913B0360, US913B0385, US913B0408, US913B0458, US913B0459, US913B0486, US913B0487, US913B0488, US913B0489, US913B0490, US913B0499, US913B0504, US913B0505, US913B0506, US913B0565, US913B0566, US913B0567, US913B0568, USO13B0040, and USO13B0041. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A problem has been detected with Philips Model EPIQ 7 Ultrasound System, software version 1.0 and 1.0.1. Using the International System Date Format may result in calculation errors in Gestational Age Gestational Age (GA) and Established Due Date (EDD).
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Medical Systems North America Inc.
- Distributor
- Worldwide Distribution-USA (nationwide) including the states of CT, HI, IL, IN, KY, MA, ME, MI, NC, OR, PA, TX, and WA, and the countries of Australia, Austria, Belgium, Chile, Denmark, Finland, France, Germany, India, Ireland, Republic of Korea, Latvia, Netherlands, New Caledonia, New Zealand, Norway, Poland, Portugal, Romania, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, and United Kingdom.
- Lot
- TRACE ITEM SERIAL NUMBERS in the UNITED STATES: US713B0171, US713B0207, US813B0017, US813B0018, US813B0019, US813B0020, US813B0022, US813B0023, US813B0024, US813B0025, US813B0122, US813B0123, US813B0125, US813B0127, US813B0128, US813B0129, US813B0131, US813B0132, US813B0134, US813B0135, US813B0233, US813B0234, US813B0235, US813B0236, US813B0237, US813B0238, US813B0239, US813B0240, US813B0423, US813B0424, US813B0425, US813B0426, US813B0427, US813B0428, US813B0429, US813B0430, US813B0431, US813B0432, US813B0433, US813B0434, US913B0033, US913B0034, US913B0035, US913B0036, US913B0037, US913B0038, US913B0039, US913B0040, US913B0057, US913B0060, US913B0061, US913B0062, US913B0063, US913B0064, US913B0065, US913B0066, US913B0106, US913B0107, US913B0108, US913B0109, US913B0110, US913B0111, US913B0112, US913B0113, US913B0144, US913B0176, US913B0201, US913B0256, US913B0257, US913B0350, US913B0363, US913B0364, US913B0365, US913B0366, US913B0367, US913B0368, US913B0369, US913B0380, US913B0384, US913B0395, US913B0396, US913B0397, US913B0398, US913B0399, US913B0400, US913B0401, US913B0402, US913B0409, US913B0441, US913B0444, US913B0445, US913B0446, US913B0447, US913B0503, US913B0522, US913B0614, US913B0615, US913B0616, US913B0618, US913B0619, and USO13B0043. TRACE ITEM SERIAL NUMBERS INTERNATIONAL: US513B0224, US513B0295, US513B0387, US513B0560, US613B0299, US713B0085, US713B0086, US713B0118, US713B0119, US713B0120, US713B0122, US713B0123, US713B0126, US713B0149, US713B0171, US713B0172, US713B0192, US713B0193, US713B0194, US713B0195, US713B0204, US713B0205, US713B0206, US713B0208, US713B0324, US713B0325, US713B0358, US713B0359, US713B0437, US713B0439, US713B0440, US713B0532, US713B0537, US713B0573, US713B0574, US713B0575, US713B0576, US713B0577, US713B0578, US713B0579, US713B0580, US713B0581, US713B0582, US713B0583, US713B0584, US713B0586, US713B0587, US713B0588, US713B0589, US713B0590, US713B0592, US713B0593, US713B0594, US713B0595, US813B0015, US813B0016, US813B0032, US813B0037, US813B0038, US813B0039, US813B0121, US813B0151, US813B0152, US813B0153, US813B0154, US813B0155, US813B0156, US813B0157, US813B0158, US813B0159, US813B0160, US813B0161, US813B0162, US813B0163, US813B0164, US813B0183, US813B0184, US813B0185, US813B0186, US813B0187, US813B0188, US813B0189, US813B0190, US813B0191, US813B0192, US813B0193, US813B0194, US813B0195, US813B0196, US813B0232, US813B0396, US813B0397, US813B0398, US813B0399, US813B0400, US813B0401, US813B0402, US813B0403, US813B0404, US813B0405, US813B0406, US813B0407, US813B0408, US813B0409, US813B0410, US813B0411, US813B0468, US813B0469, US813B0470, US813B0471, US813B0472, US813B0473, US813B0474, US813B0475, US813B0476, US813B0477, US813B0478, US813B0479, US813B0480, US913B0026, US913B0027, US913B0028, US913B0029, US913B0030, US913B0031, US913B0032, US913B0041, US913B0051, US913B0059, US913B0085, US913B0086, US913B0101, US913B0102, US913B0103, US913B0104, US913B0105, US913B0130, US913B0132, US913B0133, US913B0134, US913B0135, US913B0137, US913B0138, US913B0139, US913B0140, US913B0141, US913B0142, US913B0143, US913B0177, US913B0178, US913B0179, US913B0185, US913B0202, US913B0216, US913B0217, US913B0218, US913B0219, US913B0220, US913B0221, US913B0222, US913B0223, US913B0232, US913B0252, US913B0253, US913B0254, US913B0255, US913B0259, US913B0260, US913B0299, US913B0300, US913B0301, US913B0302, US913B0303, US913B0316, US913B0317, US913B0318, US913B0319, US913B0332, US913B0333, US913B0336, US913B0337, US913B0338, US913B0339, US913B0340, US913B0349, US913B0359, US913B0360, US913B0385, US913B0408, US913B0458, US913B0459, US913B0486, US913B0487, US913B0488, US913B0489, US913B0490, US913B0499, US913B0504, US913B0505, US913B0506, US913B0565, US913B0566, US913B0567, US913B0568, USO13B0040, and USO13B0041.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPhilips HeartStart FR3 Automated External Defibrillator (AED), models 861388 and 861389. The HeartStart FR3 isMarch 26, 2014
- FDAPhilips Model EPIQ 7 Ultrasound System. Intended for diagnostic ultrasound imaging and fluid flow analysis.March 26, 2014
- FDAPhilips Model EPIQ 5 Ultrasound System. Intended for diagnostic ultrasound imaging and fluid flow analysis.March 26, 2014
- FDAPhilips HeartStart Home and OnSite (HS1) automated external defibrillators (AEDs). Product Usage: The modelsMarch 26, 2014
- FDAGE Healthcare Discovery NMCT 670, Discovery NM 630, Optima NMCT 640, & Brivo NM 615. The GE Discovery NMJCTNovember 27, 2013
- FDASynthes Titanium Trochanteric Fixation Nail (TFN)-Helical Blade Synthes Titanium Trochanteric Fixation NailNovember 27, 2013
Questions about this recall
Was EPIQ 7 Ultrasound System, Model 795200, 795201, 795202. Intended for diagnostic ultrasound imaging and fluid flow analysis. recalled?
Yes. FDA published recall Z-0368-2014 on November 27, 2013. Reported hazard: A problem has been detected with Philips Model EPIQ 7 Ultrasound System, software version 1.0 and 1.0.1. Using the International System Date Format may result in calculation errors in Gestational Age Gestational Age (GA) and Established Due Date (EDD).
What should I do if I have this product?
Voluntary: Firm initiated · Letter