Philips HeartStart FR3 Automated External Defibrillator (AED), models 861388 and 861389. The HeartStart FR3 is a battery powered automated external defibrillator (AED) designed for use by responders who have been trained in Basic Life Support (BLS), Advanced Life Support (ALS), or another physician-authorized emergency medical response program. The HeartStart FR3 is used to treat suspected victims of ventricular fibrillation (VF), the most common cause of sudden cardiac arrest (SCA), and certain ventricular tachycardia (VTs). Both models are used with disposable defibrillator pads applied to potential victims of SCA with the following symptoms: -Unresponsiveness -Absence of normal breathing If in doubt, apply the pads. The HeartStart FR3 is intended for adults and children over 55 pounds (25 kg) or 8 years old. Both models 861388 and 861389 are also intended for children under 55 pounds (25 kg) or 8 years old when used with the optional Infant/Child Key. If the Infant/Child Key is not available, or you are uncertain of the child's age or weight, do not delay treatment.

Published March 26, 2014 · Recall initiated: February 6, 2014

Summary

FDA records medical device Class II recall Z-1225-2014, published March 26, 2014 and initiated February 6, 2014. Recalling firm: Philips Medical Systems North America Inc.; Status: Terminated; Reported quantity: 251 units total (128 in USA and 123 untils outside USA); Lot: Serial numbers of units distributed in the USA: C13A-00592, C13C-00131, C13C-00150, C13C-00270, C13C-00679, C13C-00681, C13C-00769, C13D-00048, C13D-00197, C13E-00088, C13E-00168, C13E-00234, C13E-00242, C13E-00245, C13E-00326, C13E-00340, C13E-00368, C13E-00372, C13E-00380, C13E-00382, C13E-00390, C13E-00406, C13E-00495, C13E-00507, C13E-00511, C13E-00516, C13E-00526, C13E-00537, C13F-00011, C13F-00012, C13F-00013, C13F-00022, C13F-00024, C13F-00027, C13F-00031, C13F-00032, C13F-00033, C13F-00035, C13F-00038, C13F-00043, C13F-00047, C13F-00053, C13F-00054, C13F-00059, C13F-00071, C13F-00077, C13F-00078, C13F-00102, C13F-00104, C13F-00110, C13F-00114, C13F-00118, C13F-00126, C13F-00131, C13F-00137, C13F-00138, C13F-00140, C13F-00143, C13F-00146, C13F-00148, C13F-00168, C13F-00172, C13F-00173, C13F-00174, C13F-00180, C13F-00181, C13F-00198, C13F-00202, C13F-00212, C13F-00214, C13F-00215, C13F-00217, C13F-00218, C13F-00224, C13F-00225, C13F-00227, C13F-00232, C13F-00233, C13F-00236, C13F-00238, C13F-00239, C13F-00240, C13F-00244, C13F-00246, C13F-00250, C13F-00255, C13F-00266, C13F-00273, C13F-00274, C13F-00275, C13F-00278, C13F-00284, C13F-00291, C13F-00294, C13F-00307, C13F-00309, C13F-00318, C13F-00321, C13F-00322, C13F-00323, C13F-00329, C13F-00330, C13F-00331, C13F-00332, C13F-00336, C13F-00348, C13F-00353, C13F-00468, C13F-00543, C13F-00559, C13F-00567, C13F-00578, C13F-00590, C13F-00593, C13F-00595, C13F-00599, C13F-00604, C13F-00607, C13F-00651, C13F-00778, C13G-00008, C13G-00010, C13G-00035, C13G-00056, C13G-00057, C13G-00282, C13G-00284, and C13H-00062. Serial numbers of units outside the USA: C13C-00004, C13C-00256, C13C-00305, C13C-00323, C13C-00608, C13C-00639, C13E-00009, C13E-00021, C13E-00036, C13E-00048, C13E-00057, C13E-00058, C13E-00065, C13E-00163, C13E-00250, C13E-00265, C13E-00273, C13E-00311, C13E-00323, C13E-00328, C13E-00339, C13E-00341, C13E-00345, C13E-00346, C13E-00349, C13E-00350, C13E-00351, C13E-00358, C13E-00393, C13E-00395, C13E-00404, C13E-00407, C13E-00409, C13E-00424, C13E-00437, C13E-00441, C13E-00452, C13E-00454, C13E-00455, C13E-00499, C13E-00503, C13E-00518, C13E-00527, C13E-00535, C13E-00548, C13F-00010, C13F-00280, C13F-00286, C13F-00293, C13F-00298, C13F-00302, C13F-00303, C13F-00305, C13F-00306, C13F-00316, C13F-00317, C13F-00343, C13F-00345, C13F-00346, C13F-00358, C13F-00363, C13F-00369, C13F-00371, C13F-00381, C13F-00385, C13F-00391, C13F-00400, C13F-00420, C13F-00424, C13F-00428, C13F-00436, C13F-00473, C13F-00498, C13F-00539, C13F-00565, C13F-00579, C13F-00580, C13F-00583, C13F-00586, C13F-00587, C13F-00606, C13F-00619, C13F-00627, C13F-00629, C13F-00649, C13F-00666, C13F-00668, C13F-00677, C13F-00678, C13F-00682, C13F-00685, C13F-00687, C13F-00726, C13F-00742, C13F-00746, C13F-00748, C13F-00755, C13F-00758, C13F-00760, C13F-00782, C13F-00797, C13F-00815, C13F-00870, C13F-00889, C13F-00896, C13F-00938, C13G-00026, C13G-00039, C13G-00084, C13G-00126, C13G-00132, C13G-00142, C13G-00237, C13G-00247, C13G-00293, C13G-00315, C13G-00331, C13G-00337, C13G-00345, C13G-00407, C13G-00435, C13G-00451, and C13H-00035. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Philip's is recalling HeartStart FR3 Automated External Defibrillator (AED) because the Printed Circuit Board (PCB), which contains the circuitry that supplies energy to the device, could become separated from the electrical contact points that hold the board in place. In the event the PCB becomes separated from the electrical contact points, the AED may not be ready to deliver a shock in the event of an emergency.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips Medical Systems North America Inc.
Distributor
Worldwide Distribution - US Distribution including the states of AL, CA, CO, WA DC, FL, GA, HI, KS, MD, ME, MI, MS, NC, NM, NY, PA, TX, VA, WA and WI., and the countries of Australia, Austria, Canada, France, Germany, India, Italy, Ireland, Japan, Malaysia, Netherlands, Norway, Switzerland, Taiwan and United Kingdom.
Lot
Serial numbers of units distributed in the USA: C13A-00592, C13C-00131, C13C-00150, C13C-00270, C13C-00679, C13C-00681, C13C-00769, C13D-00048, C13D-00197, C13E-00088, C13E-00168, C13E-00234, C13E-00242, C13E-00245, C13E-00326, C13E-00340, C13E-00368, C13E-00372, C13E-00380, C13E-00382, C13E-00390, C13E-00406, C13E-00495, C13E-00507, C13E-00511, C13E-00516, C13E-00526, C13E-00537, C13F-00011, C13F-00012, C13F-00013, C13F-00022, C13F-00024, C13F-00027, C13F-00031, C13F-00032, C13F-00033, C13F-00035, C13F-00038, C13F-00043, C13F-00047, C13F-00053, C13F-00054, C13F-00059, C13F-00071, C13F-00077, C13F-00078, C13F-00102, C13F-00104, C13F-00110, C13F-00114, C13F-00118, C13F-00126, C13F-00131, C13F-00137, C13F-00138, C13F-00140, C13F-00143, C13F-00146, C13F-00148, C13F-00168, C13F-00172, C13F-00173, C13F-00174, C13F-00180, C13F-00181, C13F-00198, C13F-00202, C13F-00212, C13F-00214, C13F-00215, C13F-00217, C13F-00218, C13F-00224, C13F-00225, C13F-00227, C13F-00232, C13F-00233, C13F-00236, C13F-00238, C13F-00239, C13F-00240, C13F-00244, C13F-00246, C13F-00250, C13F-00255, C13F-00266, C13F-00273, C13F-00274, C13F-00275, C13F-00278, C13F-00284, C13F-00291, C13F-00294, C13F-00307, C13F-00309, C13F-00318, C13F-00321, C13F-00322, C13F-00323, C13F-00329, C13F-00330, C13F-00331, C13F-00332, C13F-00336, C13F-00348, C13F-00353, C13F-00468, C13F-00543, C13F-00559, C13F-00567, C13F-00578, C13F-00590, C13F-00593, C13F-00595, C13F-00599, C13F-00604, C13F-00607, C13F-00651, C13F-00778, C13G-00008, C13G-00010, C13G-00035, C13G-00056, C13G-00057, C13G-00282, C13G-00284, and C13H-00062. Serial numbers of units outside the USA: C13C-00004, C13C-00256, C13C-00305, C13C-00323, C13C-00608, C13C-00639, C13E-00009, C13E-00021, C13E-00036, C13E-00048, C13E-00057, C13E-00058, C13E-00065, C13E-00163, C13E-00250, C13E-00265, C13E-00273, C13E-00311, C13E-00323, C13E-00328, C13E-00339, C13E-00341, C13E-00345, C13E-00346, C13E-00349, C13E-00350, C13E-00351, C13E-00358, C13E-00393, C13E-00395, C13E-00404, C13E-00407, C13E-00409, C13E-00424, C13E-00437, C13E-00441, C13E-00452, C13E-00454, C13E-00455, C13E-00499, C13E-00503, C13E-00518, C13E-00527, C13E-00535, C13E-00548, C13F-00010, C13F-00280, C13F-00286, C13F-00293, C13F-00298, C13F-00302, C13F-00303, C13F-00305, C13F-00306, C13F-00316, C13F-00317, C13F-00343, C13F-00345, C13F-00346, C13F-00358, C13F-00363, C13F-00369, C13F-00371, C13F-00381, C13F-00385, C13F-00391, C13F-00400, C13F-00420, C13F-00424, C13F-00428, C13F-00436, C13F-00473, C13F-00498, C13F-00539, C13F-00565, C13F-00579, C13F-00580, C13F-00583, C13F-00586, C13F-00587, C13F-00606, C13F-00619, C13F-00627, C13F-00629, C13F-00649, C13F-00666, C13F-00668, C13F-00677, C13F-00678, C13F-00682, C13F-00685, C13F-00687, C13F-00726, C13F-00742, C13F-00746, C13F-00748, C13F-00755, C13F-00758, C13F-00760, C13F-00782, C13F-00797, C13F-00815, C13F-00870, C13F-00889, C13F-00896, C13F-00938, C13G-00026, C13G-00039, C13G-00084, C13G-00126, C13G-00132, C13G-00142, C13G-00237, C13G-00247, C13G-00293, C13G-00315, C13G-00331, C13G-00337, C13G-00345, C13G-00407, C13G-00435, C13G-00451, and C13H-00035.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Philips HeartStart FR3 Automated External Defibrillator (AED), models 861388 and 861389. The HeartStart FR3 is a battery powered automated external defibrillator (AED) designed for use by responders who have been trained in Basic Life Support (BLS), Advanced Life Support (ALS), or another physician-authorized emergency medical response program. The HeartStart FR3 is used to treat suspected victims of ventricular fibrillation (VF), the most common cause of sudden cardiac arrest (SCA), and certain ventricular tachycardia (VTs). Both models are used with disposable defibrillator pads applied to potential victims of SCA with the following symptoms: -Unresponsiveness -Absence of normal breathing If in doubt, apply the pads. The HeartStart FR3 is intended for adults and children over 55 pounds (25 kg) or 8 years old. Both models 861388 and 861389 are also intended for children under 55 pounds (25 kg) or 8 years old when used with the optional Infant/Child Key. If the Infant/Child Key is not available, or you are uncertain of the child's age or weight, do not delay treatment. recalled?

Yes. FDA published recall Z-1225-2014 on March 26, 2014. Reported hazard: Philip's is recalling HeartStart FR3 Automated External Defibrillator (AED) because the Printed Circuit Board (PCB), which contains the circuitry that supplies energy to the device, could become separated from the electrical contact points that hold the board in place. In the event the PCB becomes separated from the electrical contact points, the AED may not be ready to deliver a shock in the event of an emergency.

What should I do if I have this product?

Voluntary: Firm initiated · Letter