FDAFDA · Food and drugs·Recall number Z-0369-2024
Medline manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Adult Resuscitator REF #s: CPRM1116FPM, CPRM1116FPM4, CPRM1116M, CPRM1116MC, CPRM1116PM, CPRM1116PMT, CPRM1126FM, CPRM1126FPM, CPRM1126FPWM, CPRM1126M; 2. Infant Resuscitator REF #s: CPRM3312FM, CPRM3312FM123, CPRM3312FPM, CPRM3312FPM3, CPRM3312FPMC, CPRM3312M, CPRM3312M3, CPRM3312P3M3, CPRM3312PM, CPRM3312PM2, CPRM3312PMC2, CPRM3312PMTD1, CPRM3312PMX, CPRM3312PWM, CPRM3322FM, CPRM3322M, CPRM3322PM; 3. Pediatric Resuscitator REF #s: CPRM2216FM, CPRM2216FM3, CPRM2216FPM, CPRM2216FPM3, CPRM2216FPMC, CPRM2216M, CPRM2216P3M3, CPRM2216PCM, CPRM2216PM, CPRM2216PM6, CPRM2226FM, CPRM2226FM3, CPRM2226FPM, CPRM2226FPWM, CPRM2226M; 4. Small Adult Resuscitator REF #s: CPRM4416FPMC, CPRM4416M, CPRM4426M.
Published November 29, 2023 · Recall initiated: October 11, 2023
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
What to do
Voluntary: Firm initiated · E-Mail
Recall details
- Manufacturer
- SunMed Holdings, LLC
- Distributor
- US Nationwide distribution including in the states of AR, AZ, CA¿CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MN, NC, ND, NH, NJ, NM, NV, NY, OH, SC, TN, TX, UT, VA, & WV.
- Lot
- UDI-DI:10888277628298, (Case UDI: 20888277698885, ITEM # CPRM1116FM); UDI-DI:10888277667785, (Case UDI: 20888277667782, ITEM # CPRM1116FMT); UDI-DI:10888277654815, (Case UDI: 20888277654812, ITEM # CPRM1116FPM); UDI-DI:10193489082685, (Case UDI: 20193489021544, ITEM # CPRM1116FPM4); UDI-DI:10888277628304, (Case UDI: 20888277628301, ITEM # CPRM1116M); UDI-DI:10888277698826, (Case UDI: 20888277698823, ITEM # CPRM1116MC); UDI-DI:10888277628311, (Case UDI: 20888277628318, ITEM # CPRM1116PM); UDI-DI:10888277674004, (Case UDI: 20888277674001, ITEM # CPRM1116PMT); UDI-DI:10888277629677, (Case UDI: 20888277629674, ITEM # CPRM1126FM); UDI-DI:10884389810293, (Case UDI: 20884389810290, ITEM # CPRM1126FPM); UDI-DI:10193489001945, (Case UDI: 20193489001942, ITEM # CPRM1126FPWM); UDI-DI:10888277628328, (Case UDI: 20888277628325, ITEM # CPRM1126M); UDI-DI:10888277655140, (Case UDI: 20888277655147, ITEM # CPRM2216FM); UDI-DI:10888277698857, (Case UDI: 20888277717678, ITEM # CPRM2216FM3); UDI-DI:10888277632332, (Case UDI: 20888277632339, ITEM # CPRM2216FPM); UDI-DI:10193489001150, (Case UDI: 20193489001157, ITEM # CPRM2216FPM3); UDI-DI:10193489181937, (Case UDI: 20193489181934, ITEM # CPRM2216FPMC); UDI-DI:10888277632349, (Case UDI: 20888277632346, ITEM # CPRM2216M); UDI-DI:10193489009750, (Case UDI: 20193489009757, ITEM # CPRM2216P3M3); UDI-DI:10888277669833, (Case UDI: 20888277669830, ITEM # CPRM2216PCM); UDI-DI:10888277629684, (Case UDI: 20888277629681, ITEM # CPRM2216PM); UDI-DI:10193489016000, (Case UDI: 40884389121260, ITEM # CPRM2216PM6); UDI-DI:10888277655157, (Case UDI: 20888277655154, ITEM # CPRM2226FM); UDI-DI:10888277659766, (Case UDI: 20888277659763, ITEM # CPRM2226FM3); UDI-DI:10884389811337, (Case UDI: 20884389811334, ITEM # CPRM2226FPM); UDI-DI:10193489001952, (Case UDI: 20193489001959, ITEM # CPRM2226FPWM); UDI-DI:10888277632356, (Case UDI: 20888277632353, ITEM # CPRM2226M); UDI-DI:10888277656130, (Case UDI: 20888277656137, ITEM # CPRM3312FM); UDI-DI:10888277698871, (Case UDI: 20888277698878, ITEM # CPRM3312FM123); UDI-DI:10888277654822, (Case UDI: 20888277654829, ITEM # CPRM3312FPM); UDI-DI:10193489001167, (Case UDI: 20193489001164, ITEM # CPRM3312FPM3); UDI-DI:10193489181944, (Case UDI: 20193489181941, ITEM # CPRM3312FPMC); UDI-DI:10888277632363, (Case UDI: 20888277632360, ITEM # CPRM3312M); UDI-DI:10888277698888, (Case UDI: 40888277362605, ITEM # CPRM3312M3); UDI-DI:10193489009767, (Case UDI: 20193489009764, ITEM # CPRM3312P3M3); UDI-DI:10888277718652, (Case UDI: 20888277718569, ITEM # CPRM3312PM); UDI-DI:10884389975893, (Case UDI: 20884389975890, ITEM # CPRM3312PM2); UDI-DI:10888277675193, (Case UDI: 20888277675190, ITEM # CPRM3312PMC2); UDI-DI:10888277674028, (Case UDI: 20888277674025, ITEM # CPRM3312PMTD1); UDI-DI:10884389975909, (Case UDI: 20884389975906, ITEM # CPRM3312PMX); UDI-DI:10888277667815 (Case UDI: 20888277667812, ITEM # CPRM3312PWM); UDI-DI:10080196659777, (Case UDI: 20080196659774, ITEM # CPRM3322FM); UDI-DI:10888277632370, (Case UDI: 20888277632377, ITEM # CPRM3322M); UDI-DI:10888277718562, (Case UDI: 20888277718569, ITEM # CPRM3322PM); UDI-DI:10193489181951, (Case UDI: 20193489181958, ITEM # CPRM4416FPMC); UDI-DI:10888277700499, (Case UDI: 20888277700496, ITEM # CPRM4416M); UDI-DI:10888277700512, (Case UDI: 20888277700519, ITEM # CPRM4426M). All lots manufactured from 9/21/2018 to present. Latest expiration Date 10/11/2028.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States