The EVSRF catheter is a sterile, single-use disposable medical device for endovenous radiofrequency ablation, a procedure in which a refluxing vein is thermally coagulated to permanently eliminate the vein from blood circulation. The catheter has a long shaft for insertion into a vein, a handle with start/stop button, and a connector cable. The catheter shaft has a heating element that is energized by the digiRF radiofrequency generator, which is a multi-voltage energy delivery system with touchscreen control that automatically sets the non-adjustable treatment parameters for the catheter. A button on the catheter (or a foot pedal attached to the generator) begins an automated treatment cycle of 20 seconds long at a set temperature of 120 ¿C, and the treatment stops automatically when complete. The catheter is used within a sterile operative field, while the generator remains outside the sterile operative field at all times.
Published November 20, 2024 · Recall initiated: September 4, 2024
Summary
FDA records medical device Class II recall Z-0371-2025, published November 20, 2024 and initiated September 4, 2024. Recalling firm: Bard Peripheral Vascular Inc; Status: Ongoing; Reported quantity: 222,373 catheters; Lot: Catalog Number: VC-10A2.5-6F-100 UDI does: (01)00858254006015(17)241009(10)65497423 (01)00858254006015(17)241006(10)66030181 (01)00858254006015(17)241011(10)66599182 (01)00858254006015(17)241017(10)66634308 (01)00858254006015(17)241020(10)66784768 (01)00858254006015(17)241026(10)66793709 (01)00858254006015(17)241028(10)66793710 (01)00858254006015(17)241103(10)66897727 (01)00858254006015(17)241113(10)66947733 (01)00858254006015(17)241117(10)67054619 (01)00858254006015(17)241130(10)67308611 (01)00858254006015(17)250322(10)69548248 (01)00858254006015(17)250419(10)69548249 (01)00858254006015(17)250423(10)69548250 (01)00858254006015(17)250427(10)69567357 (01)00858254006015(17)250511(10)69567358 (01)00858254006015(17)250515(10)70322501 (01)00858254006015(17)250607(10)70523716 (01)00858254006015(17)250523(10)70597994 (01)00858254006015(17)250525(10)70751594 (01)00858254006015(17)250606(10)70791660 (01)00858254006015(17)250613(10)71054985 (01)00858254006015(17)250620(10)71197454 (01)00858254006015(17)250703(10)71197455 (01)00858254006015(17)250801(10)71197574 (01)00858254006015(17)250622(10)71319092 (01)00858254006015(17)250630(10)71361455 (01)00858254006015(17)250712(10)71589682 (01)00858254006015(17)250721(10)71780480 (01)00858254006015(17)250724(10)71780481 (01)00858254006015(17)250821(10)71974718 (01)00858254006015(17)250803(10)72167973 (01)00858254006015(17)250821(10)72216631 (01)00858254006015(17)250809(10)72216758 (01)00858254006015(17)250810(10)72282293 (01)00858254006015(17)250828(10)72440542 (01)00858254006015(17)250914(10)72491306 (01)00858254006015(17)250905(10)72609982 (01)00858254006015(17)250912(10)72735944 (01)00858254006015(17)250920(10)72845915 (01)00858254006015(17)250927(10)73024457 (01)00858254006015(17)250929(10)73184835 (01)00858254006015(17)251005(10)73218082 (01)00858254006015(17)251011(10)73319890 (01)00858254006015(17)251012(10)73424182 (01)00858254006015(17)251102(10)73441733 (01)00858254006015(17)251121(10)73808592 (01)00858254006015(17)251125(10)74103378 (01)00858254006015(17)251127(10)74171115 (01)00858254006015(17)251129(10)74219123 (01)00858254006015(17)260112(10)74355356 (01)00858254006015(17)241108(10)V67114001 (01)00858254006015(17)241108(10)V67114002 (01)00858254006015(17)241108(10)V67114003 Lot Numbers: 65497423 66030181 66599182 66634308 66784768 66793709 66793710 66897727 66947733 67054619 67308611 69548248 69548249 69548250 69567357 69567358 70322501 70523716 70597994 70751594 70791660 71054985 71197454 71197455 71197574 71319092 71361455 71589682 71780480 71780481 71974718 72167973 72216631 72216758 72282293 72440542 72491306 72609982 72735944 72845915 73024457 73184835 73218082 73319890 73424182 73441733 73808592 74103378 74171115 74219123 74355356 V67114001 V67114002 V67114003 Catalog Number: VC-10A2.5-6F-60 UDI codes: (01)00858254006008(17)241013(10)66186400 (01)00858254006008(17)241024(10)66793696 (01)00858254006008(17)241102(10)66793697 (01)00858254006008(17)241107(10)66793698 (01)00858254006008(17)241021(10)66802218 (01)00858254006008(17)241031(10)66863470 (01)00858254006008(17)241110(10)66981017 (01)00858254006008(17)241120(10)67017611 (01)00858254006008(17)241125(10)67105253 (01)00858254006008(17)241115(10)67190839 (01)00858254006008(17)241123(10)67241133 (01)00858254006008(17)241128(10)67429950 (01)00858254006008(17)241204(10)67429951 (01)00858254006008(17)241204(10)67429952 (01)00858254006008(17)250410(10)69548196 (01)00858254006008(17)250327(10)69548197 (01)00858254006008(17)250416(10)69548198 (01)00858254006008(17)250331(10)69659895 (01)00858254006008(17)250502(10)69659896 (01)00858254006008(17)250505(10)69748757 (01)00858254006008(17)250509(10)69928277 (01)00858254006008(17)250601(10)70414634 (01)00858254006008(17)250517(10)70487066 (01)00858254006008(17)250519(10)70634659 (01)00858254006008(17)250529(10)70673130 (01)00858254006008(17)250604(10)70868426 (01)00858254006008(17)250608(10)70933358 (01)00858254006008(17)250612(10)70994254 (01)00858254006008(17)250614(10)71075353 (01)0. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The Venclose" RF Ablation Catheters may have been manufactured with internal wiring error. The red and yellow signal wires were soldered on swapped ports of the catheter PCB. This error in wire configuration would lead to power being applied through the proximal 3/4th of the heating coil, bypassing the thermocouple, which will then cause the generator to overdrive the catheter in an attempt to reach desired temperature set point. The wire attachment error causes the device to reach temperatures exceeding the levels intended for therapeutic efficacy, while the console is falsely indicating that temperature has not yet been reached, thus increasing the risk of harm due to excessive temperature.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Bard Peripheral Vascular Inc
- Distributor
- Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI and WV. The countries of Australia, Singapore, Austria, Belgium, Czechia, Denmark, France, Germany, Greece, Ireland, Italy, Poland, Slovenia Spain, Sweden, Switzerland, United Kingdom, United Arab Emirates, and South Africa.
- Lot
- Catalog Number: VC-10A2.5-6F-100 UDI does: (01)00858254006015(17)241009(10)65497423 (01)00858254006015(17)241006(10)66030181 (01)00858254006015(17)241011(10)66599182 (01)00858254006015(17)241017(10)66634308 (01)00858254006015(17)241020(10)66784768 (01)00858254006015(17)241026(10)66793709 (01)00858254006015(17)241028(10)66793710 (01)00858254006015(17)241103(10)66897727 (01)00858254006015(17)241113(10)66947733 (01)00858254006015(17)241117(10)67054619 (01)00858254006015(17)241130(10)67308611 (01)00858254006015(17)250322(10)69548248 (01)00858254006015(17)250419(10)69548249 (01)00858254006015(17)250423(10)69548250 (01)00858254006015(17)250427(10)69567357 (01)00858254006015(17)250511(10)69567358 (01)00858254006015(17)250515(10)70322501 (01)00858254006015(17)250607(10)70523716 (01)00858254006015(17)250523(10)70597994 (01)00858254006015(17)250525(10)70751594 (01)00858254006015(17)250606(10)70791660 (01)00858254006015(17)250613(10)71054985 (01)00858254006015(17)250620(10)71197454 (01)00858254006015(17)250703(10)71197455 (01)00858254006015(17)250801(10)71197574 (01)00858254006015(17)250622(10)71319092 (01)00858254006015(17)250630(10)71361455 (01)00858254006015(17)250712(10)71589682 (01)00858254006015(17)250721(10)71780480 (01)00858254006015(17)250724(10)71780481 (01)00858254006015(17)250821(10)71974718 (01)00858254006015(17)250803(10)72167973 (01)00858254006015(17)250821(10)72216631 (01)00858254006015(17)250809(10)72216758 (01)00858254006015(17)250810(10)72282293 (01)00858254006015(17)250828(10)72440542 (01)00858254006015(17)250914(10)72491306 (01)00858254006015(17)250905(10)72609982 (01)00858254006015(17)250912(10)72735944 (01)00858254006015(17)250920(10)72845915 (01)00858254006015(17)250927(10)73024457 (01)00858254006015(17)250929(10)73184835 (01)00858254006015(17)251005(10)73218082 (01)00858254006015(17)251011(10)73319890 (01)00858254006015(17)251012(10)73424182 (01)00858254006015(17)251102(10)73441733 (01)00858254006015(17)251121(10)73808592 (01)00858254006015(17)251125(10)74103378 (01)00858254006015(17)251127(10)74171115 (01)00858254006015(17)251129(10)74219123 (01)00858254006015(17)260112(10)74355356 (01)00858254006015(17)241108(10)V67114001 (01)00858254006015(17)241108(10)V67114002 (01)00858254006015(17)241108(10)V67114003 Lot Numbers: 65497423 66030181 66599182 66634308 66784768 66793709 66793710 66897727 66947733 67054619 67308611 69548248 69548249 69548250 69567357 69567358 70322501 70523716 70597994 70751594 70791660 71054985 71197454 71197455 71197574 71319092 71361455 71589682 71780480 71780481 71974718 72167973 72216631 72216758 72282293 72440542 72491306 72609982 72735944 72845915 73024457 73184835 73218082 73319890 73424182 73441733 73808592 74103378 74171115 74219123 74355356 V67114001 V67114002 V67114003 Catalog Number: VC-10A2.5-6F-60 UDI codes: (01)00858254006008(17)241013(10)66186400 (01)00858254006008(17)241024(10)66793696 (01)00858254006008(17)241102(10)66793697 (01)00858254006008(17)241107(10)66793698 (01)00858254006008(17)241021(10)66802218 (01)00858254006008(17)241031(10)66863470 (01)00858254006008(17)241110(10)66981017 (01)00858254006008(17)241120(10)67017611 (01)00858254006008(17)241125(10)67105253 (01)00858254006008(17)241115(10)67190839 (01)00858254006008(17)241123(10)67241133 (01)00858254006008(17)241128(10)67429950 (01)00858254006008(17)241204(10)67429951 (01)00858254006008(17)241204(10)67429952 (01)00858254006008(17)250410(10)69548196 (01)00858254006008(17)250327(10)69548197 (01)00858254006008(17)250416(10)69548198 (01)00858254006008(17)250331(10)69659895 (01)00858254006008(17)250502(10)69659896 (01)00858254006008(17)250505(10)69748757 (01)00858254006008(17)250509(10)69928277 (01)00858254006008(17)250601(10)70414634 (01)00858254006008(17)250517(10)70487066 (01)00858254006008(17)250519(10)70634659 (01)00858254006008(17)250529(10)70673130 (01)00858254006008(17)250604(10)70868426 (01)00858254006008(17)250608(10)70933358 (01)00858254006008(17)250612(10)70994254 (01)00858254006008(17)250614(10)71075353 (01)0
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAEnCor EnCompass Breast Biopsy and Tissue Removal System Console (model number: ENCPCONSOLE) and Driver (modelOctober 7, 2026
- FDABroviac 4.2 Fr Single-Lumen CV Catheter Cutdown Tray with SureCuff Tissue Ingrowth Cuff Cat 60060February 25, 2026
- FDABard Marquee Disposable Core Biopsy Instrument Kit: BARD MARQUEE KIT 12G X 10 CM, REF: MQK1210; BARD MARQUEENovember 27, 2024
- FDAStryker Neptune SafeAir Smoke Evacuation Pencil Coated Push-Button, Bulk Non-sterile - Designed for generalNovember 20, 2024
- FDAMedtronic Recharger Kits, RS6230, RS7230, and RS5220A, containing components WR9230 and WR9220A Recharger forNovember 20, 2024
- FDABARD MAGNUM Disposable Core Tissue Biopsy Needle REF: MN1410 MN 1420 MN1610 MN1616 MN1820 MN2010 The MagnumOctober 2, 2024
Questions about this recall
Was The EVSRF catheter is a sterile, single-use disposable medical device for endovenous radiofrequency ablation, a procedure in which a refluxing vein is thermally coagulated to permanently eliminate the vein from blood circulation. The catheter has a long shaft for insertion into a vein, a handle with start/stop button, and a connector cable. The catheter shaft has a heating element that is energized by the digiRF radiofrequency generator, which is a multi-voltage energy delivery system with touchscreen control that automatically sets the non-adjustable treatment parameters for the catheter. A button on the catheter (or a foot pedal attached to the generator) begins an automated treatment cycle of 20 seconds long at a set temperature of 120 ¿C, and the treatment stops automatically when complete. The catheter is used within a sterile operative field, while the generator remains outside the sterile operative field at all times. recalled?
Yes. FDA published recall Z-0371-2025 on November 20, 2024. Reported hazard: The Venclose" RF Ablation Catheters may have been manufactured with internal wiring error. The red and yellow signal wires were soldered on swapped ports of the catheter PCB. This error in wire configuration would lead to power being applied through the proximal 3/4th of the heating coil, bypassing the thermocouple, which will then cause the generator to overdrive the catheter in an attempt to reach desired temperature set point. The wire attachment error causes the device to reach temperatures exceeding the levels intended for therapeutic efficacy, while the console is falsely indicating that temperature has not yet been reached, thus increasing the risk of harm due to excessive temperature.
What should I do if I have this product?
Voluntary: Firm initiated · Letter