Back to search
FDAFDA · Food and drugs·Recall number Z-3206-2024

BARD MAGNUM Disposable Core Tissue Biopsy Needle REF: MN1410 MN 1420 MN1610 MN1616 MN1820 MN2010 The Magnum Biopsy System (instrument and needles) is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, breast, spleen, lymph nodes and various soft tissue tumors.

Published October 2, 2024 · Recall initiated: August 2, 2024

Summary

FDA records medical device Class II recall Z-3206-2024, published October 2, 2024 and initiated August 2, 2024. Recalling firm: Bard Peripheral Vascular Inc; Status: Ongoing; Reported quantity: 60 (U.S.) and 5,160 (O.U.S.); Lot: Product Name: Magnum 14g x 10cm Catalog Number: MN1410 Lot Numbers / UDI code: REJP3295 / (01)00801741084249(17)270131(10)REJP3295 REJP2228 / (01)00801741084249(17)270131(10)REJP2228 REJN2583 / (01)00801741084249(17)261231(10)REJN2583 REJQ2851 / (01)00801741084249(17)270228(10)REJQ2851 REJQ0831 / (01)00801741084249(17)270228(10)REJQ0831 Product Name: Magnum 14g x 20cm Catalog Number: MN1420 Lot Numbers / UDI code: REHX2867 / (01)00801741084270(17)260930(10)REHX2867 Product Name: Magnum 16g x 10cm Catalog Number: MN1610 Lot Numbers / UDI code: REJN0587 / (01)00801741084287(17)261231(10)REJN0587 Product Name: Magnum 16g x 16cm Catalog Number: MN1616 Lot Numbers / UDI code: REJP3289 / (01)00801741084300(17)270131(10)REJP3289 Product Name: Magnum 18g x 20cm Catalog Number: MN1820 Lot Numbers / UDI code: REJS1418 / (01)00801741084355(17)270430(10)REJS1418 Product Name: Magnum 20g x 10cm Catalog Number: MN2010 Lot Numbers / UDI code: REJQ1160 / (01)00801741084386(17)270228(10)REJQ1160. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Due to possible packaging damage (indentations/cuts) cause by manufacturing process. As a result of the packaging damage, may cause the sterile barrier to be compromised.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Bard Peripheral Vascular Inc
Distributor
Worldwide - U.S. Nationwide distribution in the Puerto Rico and the countries of Canada, Pakistan, Peru, India, Malaysia, China, Brazil, Mexico, Japan, Chile, Australia.
Lot
Product Name: Magnum 14g x 10cm Catalog Number: MN1410 Lot Numbers / UDI code: REJP3295 / (01)00801741084249(17)270131(10)REJP3295 REJP2228 / (01)00801741084249(17)270131(10)REJP2228 REJN2583 / (01)00801741084249(17)261231(10)REJN2583 REJQ2851 / (01)00801741084249(17)270228(10)REJQ2851 REJQ0831 / (01)00801741084249(17)270228(10)REJQ0831 Product Name: Magnum 14g x 20cm Catalog Number: MN1420 Lot Numbers / UDI code: REHX2867 / (01)00801741084270(17)260930(10)REHX2867 Product Name: Magnum 16g x 10cm Catalog Number: MN1610 Lot Numbers / UDI code: REJN0587 / (01)00801741084287(17)261231(10)REJN0587 Product Name: Magnum 16g x 16cm Catalog Number: MN1616 Lot Numbers / UDI code: REJP3289 / (01)00801741084300(17)270131(10)REJP3289 Product Name: Magnum 18g x 20cm Catalog Number: MN1820 Lot Numbers / UDI code: REJS1418 / (01)00801741084355(17)270430(10)REJS1418 Product Name: Magnum 20g x 10cm Catalog Number: MN2010 Lot Numbers / UDI code: REJQ1160 / (01)00801741084386(17)270228(10)REJQ1160
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was BARD MAGNUM Disposable Core Tissue Biopsy Needle REF: MN1410 MN 1420 MN1610 MN1616 MN1820 MN2010 The Magnum Biopsy System (instrument and needles) is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, breast, spleen, lymph nodes and various soft tissue tumors. recalled?

Yes. FDA published recall Z-3206-2024 on October 2, 2024. Reported hazard: Due to possible packaging damage (indentations/cuts) cause by manufacturing process. As a result of the packaging damage, may cause the sterile barrier to be compromised.

What should I do if I have this product?

Voluntary: Firm initiated · Letter