FDAFDA · Food and drugs·Recall number Z-0379-2022
EVIS EXERA II DUODENOVIDEOSCOPE OLYMPUS TJF-Q180V Model Number: TJF-Q180V
Published December 22, 2021 · Recall initiated: November 2, 2021
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Use of a TJF-Q180V with adhesive deterioration or other damages may pose a risk of endoscope contamination due to ineffective reprocessing or fluid invasion; contaminated endoscope can present an infection risk to patients.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Olympus Corporation of the Americas
- Distributor
- US Nationwide distribution.
- Lot
- All serial numbers. UDI: 04953170339967
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States