Back to search
FDAFDA · Food and drugs·Recall number Z-0379-2022

EVIS EXERA II DUODENOVIDEOSCOPE OLYMPUS TJF-Q180V Model Number: TJF-Q180V

Published December 22, 2021 · Recall initiated: November 2, 2021

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Use of a TJF-Q180V with adhesive deterioration or other damages may pose a risk of endoscope contamination due to ineffective reprocessing or fluid invasion; contaminated endoscope can present an infection risk to patients.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Olympus Corporation of the Americas
Distributor
US Nationwide distribution.
Lot
All serial numbers. UDI: 04953170339967
Risk classification
Class II
Area (FDA)
device
Country
United States