Back to search
FDAFDA · Food and drugs·Recall number Z-0379-2022

EVIS EXERA II DUODENOVIDEOSCOPE OLYMPUS TJF-Q180V Model Number: TJF-Q180V

Published December 22, 2021 · Recall initiated: November 2, 2021

Summary

FDA records medical device Class II recall Z-0379-2022, published December 22, 2021 and initiated November 2, 2021. Recalling firm: Olympus Corporation of the Americas; Status: Ongoing; Reported quantity: 14,447 units; Lot: All serial numbers. UDI: 04953170339967. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Use of a TJF-Q180V with adhesive deterioration or other damages may pose a risk of endoscope contamination due to ineffective reprocessing or fluid invasion; contaminated endoscope can present an infection risk to patients.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Olympus Corporation of the Americas
Distributor
US Nationwide distribution.
Lot
All serial numbers. UDI: 04953170339967
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was EVIS EXERA II DUODENOVIDEOSCOPE OLYMPUS TJF-Q180V Model Number: TJF-Q180V recalled?

Yes. FDA published recall Z-0379-2022 on December 22, 2021. Reported hazard: Use of a TJF-Q180V with adhesive deterioration or other damages may pose a risk of endoscope contamination due to ineffective reprocessing or fluid invasion; contaminated endoscope can present an infection risk to patients.

What should I do if I have this product?

Voluntary: Firm initiated · Letter