Large Buttress Compression Nut For Blade Guide Sleeve TFN The Synthes Large Buttress Compression Nut For Blade Guide Sleeve TFN is intended to treat stable and unstable pertrochanteric fractures, intertrochanteric fractures, basal neck fractures, and combinations there of.
Published December 4, 2013 · Recall initiated: July 31, 2013
Summary
FDA records medical device Class II recall Z-0387-2014, published December 4, 2013 and initiated July 31, 2013. Recalling firm: Synthes USA HQ, Inc.; Status: Terminated; Reported quantity: 135; Lot: Part # Lot # SD357.371 5734551 6083065 6538811 6631975 6703191 6703190 6841538. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The Large Buttress Compression Nut For Blade Guide Sleeve TFN was recalled because of complaints that the compression nut did not thread completely onto the blade guide sleeve and became stuck.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Synthes USA HQ, Inc.
- Distributor
- US Distribution including the states of TX, PA, NY, GA, CO, IN, AI, AR, IL, MI, TN, CA, AZ, MS and VA.
- Lot
- Part # Lot # SD357.371 5734551 6083065 6538811 6631975 6703191 6703190 6841538
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASynthes Hand Switch - Electric Pen Drive, Part Number 05.001.012 The Synthes Electric Pen Drive is indicatedJanuary 8, 2014
- FDASynthes TomoFix Plates The Synthes J5606-C TomoFix Osteotomy System is used for open and closed wedgeDecember 18, 2013
- FDASynthes Trochanteric Fixation Nail As part of the Synthes Trochanteric Fixation Nail (TFN) System, the TFNDecember 18, 2013
- FDASynthes Soft Tissue Retractor Small Extendible. Offset blade facilitates easy preparation of the epipereostealDecember 11, 2013
- FDASingle Analyte Urine DAU Control (5 m L Norbuprenorphine, Level II control For In Vitro Diagnostic Use Only.December 4, 2013
- FDASynthes Tensioning Device The Synthes Osteogenesis System, MR Conditional is intended for fracture fixationDecember 4, 2013
Questions about this recall
Was Large Buttress Compression Nut For Blade Guide Sleeve TFN The Synthes Large Buttress Compression Nut For Blade Guide Sleeve TFN is intended to treat stable and unstable pertrochanteric fractures, intertrochanteric fractures, basal neck fractures, and combinations there of. recalled?
Yes. FDA published recall Z-0387-2014 on December 4, 2013. Reported hazard: The Large Buttress Compression Nut For Blade Guide Sleeve TFN was recalled because of complaints that the compression nut did not thread completely onto the blade guide sleeve and became stuck.
What should I do if I have this product?
Voluntary: Firm initiated · Letter