LANX Timberline MPF Lateral Modular Plate Fixation System Straight Split Tip Driver - Short ("Driver"). Model Number 8630-0107. The Timberline MPF device is an intervertebral body fusion device consisting of a PEEK intervertebral spacer, titanium plate, and screws. The PEEK spacer has a generally rounded shape with various heights and footprints and has a hollowed out central area to accommodate autogenous bone graft. The upper and lower surfaces have a series of transverse grooves formed to improve stability and fixation once the device is inserted. The titanium plate has holes for receiving bone screws and a central hole for receiving a cover plate to prevent screw back-out. The Timberline MPF system is available in a variety of sizes and configurations to approximate anatomical variation in different vertebral levels and/or patient anatomy. The Timberline MPF system is provided nonsterile. The Driver is used with the Timberline MPF system to drive titanium screws through the plate and into the vertebrae thus locking plate to the vertebrae.
Published December 4, 2013 · Recall initiated: November 4, 2013
Summary
FDA records medical device Class II recall Z-0400-2014, published December 4, 2013 and initiated November 4, 2013. Recalling firm: Lanx, Inc.; Status: Terminated; Reported quantity: 25 units; Lot: Lot No: TU00090, TU00157 and TU00242. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Lanx, Inc. is recalling various lots of Timberline MPF Straight Split Tip Driver - Short ("Driver") because the width of the most distal end from the Driver handle (the tip of the hex portion of the Driver) is oversized.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Lanx, Inc.
- Distributor
- Nationwide Distribution including AZ, PA, WI, NY, TX, PR, and TN.
- Lot
- Lot No: TU00090, TU00157 and TU00242
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMedtronic Strata II Shunt Assembly Kit, Small, Catalog No. 27848. Strata II Shunt Assemblies are designed toDecember 4, 2013
- FDASynthes 4.5MM TI VA-LCP CRVD Condylar Plate/6H/159MM/RT-STER Synthes 4.5MM TI VA-LCP CRVD CondylarDecember 4, 2013
- FDAMedtronic Delta Shunt Assembly Kit with Bioglide Small, Performance Level 1.0, Catalog No. 276201. Delta ShuntDecember 4, 2013
- FDASTERRAD 100NX, Product Code: 10104 The STERRAD Steri lization System is a low temperature general purposeDecember 4, 2013
- FDACSF-Unitized Shunt, Contoured Regular, Low Pressure, Catalog No. 46022; CSF-Unitized Shunt, Contoured Regular,December 4, 2013
- FDASTERRAD NX, Product Code: 10033 The STERRAD Steri lization System is a low temperature general purposeDecember 4, 2013
Questions about this recall
Was LANX Timberline MPF Lateral Modular Plate Fixation System Straight Split Tip Driver - Short ("Driver"). Model Number 8630-0107. The Timberline MPF device is an intervertebral body fusion device consisting of a PEEK intervertebral spacer, titanium plate, and screws. The PEEK spacer has a generally rounded shape with various heights and footprints and has a hollowed out central area to accommodate autogenous bone graft. The upper and lower surfaces have a series of transverse grooves formed to improve stability and fixation once the device is inserted. The titanium plate has holes for receiving bone screws and a central hole for receiving a cover plate to prevent screw back-out. The Timberline MPF system is available in a variety of sizes and configurations to approximate anatomical variation in different vertebral levels and/or patient anatomy. The Timberline MPF system is provided nonsterile. The Driver is used with the Timberline MPF system to drive titanium screws through the plate and into the vertebrae thus locking plate to the vertebrae. recalled?
Yes. FDA published recall Z-0400-2014 on December 4, 2013. Reported hazard: Lanx, Inc. is recalling various lots of Timberline MPF Straight Split Tip Driver - Short ("Driver") because the width of the most distal end from the Driver handle (the tip of the hex portion of the Driver) is oversized.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit