Back to search
FDAFDA · Food and drugs·Recall number Z-0403-2016

Olympus ENDOEYE HD II Video Telescopes. Models WA50040A ENDOEYE HD II, 10 mm, 0 degree, WA50042A ENDOEYE HD II 10 mm, 30 degree, W7100603 Repair cable 10 mm 0/30 degree. The ENDOEYE video Telescopes are used with other supporting equipment for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including the female reproductive organs.

Published December 16, 2015 · Recall initiated: November 17, 2015

Summary

FDA records medical device Class II recall Z-0403-2016, published December 16, 2015 and initiated November 17, 2015. Recalling firm: Olympus Corporation of the Americas; Status: Terminated; Reported quantity: 182; Lot: Serial numbers 622483, 622523, 622553, 622566, 622568, 622569, 622573, 622576, 622614, 622615, 622616, 622617, 622618, 622619, 622620, 622632, 622643, 622644, 622645, 622646, 622647, 622654, 622655, 622656, 622657, 622658, 622664, 622666, 622667, 622687, 622688, 622693, 622704, 622705, 622706, 622710, 622711, 622712, 622718, 622719, 622720, 622722, 622723, 622750, 622753, 622764, 622766, 622767, 622769, 622771, 622773, 622775, 622777, 622778, 622785, 622789, 622795, 622797, 622798, 622801, 622806, 623624, 623626, 623629, 623637, 623638, 623639, 623641, 623661, 623669, 623671, 623672, 623673, 623677, 623687, 623688, 623705, 623709, 623713, 623719, 623720, 623722, 623723, 623729, 623730, 623731, 623735, 623736, 623737, 623738, 623739, 623740, 623741, 623742, 623743, 623747, 623749, 623750, 623751, 623752, 623776, 623778, 623784, 623786, 623787, 623788, 623799, 623800, 623801, 623802, 624004, 624005, 624006, 624008, 624011, 624012, 624016, 624017, 624018, 624019, 624024, 624028, 624030, 624041, 624042, 624046, 624054, 624055, 624056, 624057, 624064, 624066, 624070, 624071, 624073, 624076, 624077, 624080, 624082, 624089, 624099, 624100, 624101, 624104, 624107, 624109, 624110, 624112, 624114, 624117, 624119, 624121, 624151, 624158, 624162, 624167, 624170, 624618, 624622, 624643, 624678, 624679, 624682, 624683, 624720, 622539, 622540, 622570, 622572, 623734, 623748, 624739, and 624741. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A damaged temperature sensor at the ENDOEYE tip caused the distal end to become abnormally hot. Excessive heating of the ENDOEYE distal end could result in patient or user injury.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Olympus Corporation of the Americas
Distributor
US Nationwide Distribution including these states: NE, IL, FL, TX, OH, PA, IN, CA, NY, AL, ND, GA, AK, MI, WA, MS, NJ, IA, AZ, and RI.
Lot
Serial numbers 622483, 622523, 622553, 622566, 622568, 622569, 622573, 622576, 622614, 622615, 622616, 622617, 622618, 622619, 622620, 622632, 622643, 622644, 622645, 622646, 622647, 622654, 622655, 622656, 622657, 622658, 622664, 622666, 622667, 622687, 622688, 622693, 622704, 622705, 622706, 622710, 622711, 622712, 622718, 622719, 622720, 622722, 622723, 622750, 622753, 622764, 622766, 622767, 622769, 622771, 622773, 622775, 622777, 622778, 622785, 622789, 622795, 622797, 622798, 622801, 622806, 623624, 623626, 623629, 623637, 623638, 623639, 623641, 623661, 623669, 623671, 623672, 623673, 623677, 623687, 623688, 623705, 623709, 623713, 623719, 623720, 623722, 623723, 623729, 623730, 623731, 623735, 623736, 623737, 623738, 623739, 623740, 623741, 623742, 623743, 623747, 623749, 623750, 623751, 623752, 623776, 623778, 623784, 623786, 623787, 623788, 623799, 623800, 623801, 623802, 624004, 624005, 624006, 624008, 624011, 624012, 624016, 624017, 624018, 624019, 624024, 624028, 624030, 624041, 624042, 624046, 624054, 624055, 624056, 624057, 624064, 624066, 624070, 624071, 624073, 624076, 624077, 624080, 624082, 624089, 624099, 624100, 624101, 624104, 624107, 624109, 624110, 624112, 624114, 624117, 624119, 624121, 624151, 624158, 624162, 624167, 624170, 624618, 624622, 624643, 624678, 624679, 624682, 624683, 624720, 622539, 622540, 622570, 622572, 623734, 623748, 624739, and 624741.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Olympus ENDOEYE HD II Video Telescopes. Models WA50040A ENDOEYE HD II, 10 mm, 0 degree, WA50042A ENDOEYE HD II 10 mm, 30 degree, W7100603 Repair cable 10 mm 0/30 degree. The ENDOEYE video Telescopes are used with other supporting equipment for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including the female reproductive organs. recalled?

Yes. FDA published recall Z-0403-2016 on December 16, 2015. Reported hazard: A damaged temperature sensor at the ENDOEYE tip caused the distal end to become abnormally hot. Excessive heating of the ENDOEYE distal end could result in patient or user injury.

What should I do if I have this product?

Voluntary: Firm initiated · Letter