Guardia Access Nano Embryo Transfer Catheter is use for transferring IVF embryos into the uterine cavity.
Published November 27, 2019 · Recall initiated: October 3, 2019
Summary
FDA records medical device Class II recall Z-0415-2020, published November 27, 2019 and initiated October 3, 2019. Recalling firm: Cook Inc.; Status: Terminated; Reported quantity: 202; Lot: Lot Number 9502915. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Bent transfer catheter tips may cause difficulty or inability in inserting the transfer catheter into the guiding catheter. Potential adverse events that may occur if an affected product is used include increased procedural time, the need for a repeat embryo transfer procedure, or the need for a patient to undergo an additional IVF cycle.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Cook Inc.
- Distributor
- No US distribution. The products were distributed to the following foreign countries: Belgium, Denmark, Germany, Italy, Netherlands, Norway, Poland, Republic of Macedonia, Slovenia, Spain, and Switzerland.
- Lot
- Lot Number 9502915
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFlexor Check-Flo Introducer - Raabe Modification, Introducer Set, RPN KCFW-5.0-38-90-RB-RAABE, GPN G13069 TheNovember 10, 2021
- FDAFlexor Check-Flo Introducer, RPN KCFW-8.0-35-55-RB-HFANL0-HC, Order Number G56233 Flexor Introducers andDecember 30, 2020
- FDAFlexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL0-HC, Order Number G56231 Flexor Introducers andDecember 30, 2020
- FDAFlexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL1-HC, Order Number G56236 Flexor Introducers andDecember 30, 2020
- FDARandox Liquid Cardiac Controls Catalogue Numbers CQ5053.November 27, 2019
- FDABlanketrol III Hyper-Hypothermia System- body hypothermia system temperature management Model 233 - ProductNovember 27, 2019
Questions about this recall
Was Guardia Access Nano Embryo Transfer Catheter is use for transferring IVF embryos into the uterine cavity. recalled?
Yes. FDA published recall Z-0415-2020 on November 27, 2019. Reported hazard: Bent transfer catheter tips may cause difficulty or inability in inserting the transfer catheter into the guiding catheter. Potential adverse events that may occur if an affected product is used include increased procedural time, the need for a repeat embryo transfer procedure, or the need for a patient to undergo an additional IVF cycle.
What should I do if I have this product?
Voluntary: Firm initiated · Letter