Back to search
FDAFDA · Food and drugs·Recall number Z-0425-2014

13mm addEASE(TM) Binary Connector with 17 Ga. Needle for use with PAB(R) Containers, BC1300 Product is intended for fluid transfer between a Partial Additive Bag (PAB(R)) container and a drug vial.

Published December 4, 2013 · Recall initiated: April 4, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Possible solution leakage at the interface of the addEASE(TM) Binary Connector (Product Codes BC2000 and BC1300) and the PAB(R) bag. Based on extensive root cause investigation, it has been determined that the leakage is the result of a misaligned cannula (needle) on the addEASE(TM) Binary Connector, which pierces the PAB(R) bag stopper to access the solution. The misaligned cannula may result i

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
B. Braun Medical, Inc.
Distributor
Nationwide Distribution
Lot
Material BC1300, Lot number 61286479.
Risk classification
Class II
Area (FDA)
device
Country
United States