13mm addEASE(TM) Binary Connector with 17 Ga. Needle for use with PAB(R) Containers, BC1300 Product is intended for fluid transfer between a Partial Additive Bag (PAB(R)) container and a drug vial.

Published December 4, 2013 · Recall initiated: April 4, 2013

Summary

FDA records medical device Class II recall Z-0425-2014, published December 4, 2013 and initiated April 4, 2013. Recalling firm: B. Braun Medical, Inc.; Status: Terminated; Reported quantity: 4,000 units; Lot: Material BC1300, Lot number 61286479. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Possible solution leakage at the interface of the addEASE(TM) Binary Connector (Product Codes BC2000 and BC1300) and the PAB(R) bag. Based on extensive root cause investigation, it has been determined that the leakage is the result of a misaligned cannula (needle) on the addEASE(TM) Binary Connector, which pierces the PAB(R) bag stopper to access the solution. The misaligned cannula may result i

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
B. Braun Medical, Inc.
Distributor
Nationwide Distribution
Lot
Material BC1300, Lot number 61286479.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was 13mm addEASE(TM) Binary Connector with 17 Ga. Needle for use with PAB(R) Containers, BC1300 Product is intended for fluid transfer between a Partial Additive Bag (PAB(R)) container and a drug vial. recalled?

Yes. FDA published recall Z-0425-2014 on December 4, 2013. Reported hazard: Possible solution leakage at the interface of the addEASE(TM) Binary Connector (Product Codes BC2000 and BC1300) and the PAB(R) bag. Based on extensive root cause investigation, it has been determined that the leakage is the result of a misaligned cannula (needle) on the addEASE(TM) Binary Connector, which pierces the PAB(R) bag stopper to access the solution. The misaligned cannula may result i

What should I do if I have this product?

Voluntary: Firm initiated · Letter