AET36 ANESTH EXT SET - Anesthesia Extension Set - Normally Closed Dual Manifold, Four-way Stopcock, Model/Catalog Number: 472036

Published September 25, 2024 · Recall initiated: August 8, 2024

Summary

FDA records medical device Class II recall Z-3144-2024, published September 25, 2024 and initiated August 8, 2024. Recalling firm: B. Braun Medical, Inc.; Status: Ongoing; Reported quantity: 19050 units; Lot: Model No 472036; UDI-DI 04046964184659; Lots 0061898815, 0061905158, 00VL872690, 00VL882317 and 00VL882319. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Potential for product leakage from the Spin Lock Connector on the distal end of the device due to the migration of excess solvent during the assembly process.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
B. Braun Medical, Inc.
Distributor
US Nationwide distribution.
Lot
Model No 472036; UDI-DI 04046964184659; Lots 0061898815, 0061905158, 00VL872690, 00VL882317 and 00VL882319
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was AET36 ANESTH EXT SET - Anesthesia Extension Set - Normally Closed Dual Manifold, Four-way Stopcock, Model/Catalog Number: 472036 recalled?

Yes. FDA published recall Z-3144-2024 on September 25, 2024. Reported hazard: Potential for product leakage from the Spin Lock Connector on the distal end of the device due to the migration of excess solvent during the assembly process.

What should I do if I have this product?

Voluntary: Firm initiated · Letter